T703 Calibration and Verification Teledyne API T703 Calibrator Operation Manual

142

8.3. T703 GAS PRESSURE SENSOR CALIBRATION

The T703 Calibrator has two sensors that monitor the pressure of the gases flowing through the instrument. The

data collected by these sensors is used to compensate the final concentration calculations for changes in

atmospheric pressure and is stored in the CPU’s memory as test functions:

Table 8-1: T703 Pressure Sensors

SENSOR ASSOCIATED
TEST FUNCTION UNITS PRESSURE MONITOR
MEASUREMENT POINT
Regulator Pressure Sensor
REG PRESSURE PSIG Capped fitting on backside of regulator
assembly. See Figure 8-4
Photometer Sample Gas
Pressure Sensor PHOTO SPRESS IN-HG-A
Use monitor to measure ambient
atmospheric pressure at the calibrator’s
location.

8.3.1.1. Gas Pressure Sensor Calibration Set Up

The procedures described in this section require an independent, calibrated pressure meter/monitor be attached

at the following location.

PHOTOMETER
PRESSURE SENSOR
O3GEN / PHOTOMETER
PRESSURE / FLOW SENSOR PCA
O3GAS INPUT
PRESSURE SENSOR
O3FLOW
SENSOR
O3Generator Assembly
O3
GENERATOR
Flow Control
(100 cm3)
REF/MEAS
Valve
On Back Panel
Chassis
GAS OUTPUT MANIFOLD
PHOTOMETER
OUTLET
TO ANALYZER
VENT
TO ANALYZER
DRY AIR
IN
ZERO AIR
IN
PHOTOMETER
INLET
EXHAUST
PHOTOMETER
ZERO OUT
PHOTOMETER
ZERO IN
PHOTOMETER BENCH
PUMP
Flow Control
(800 cm3)
INTERNAL
VENT
Flow Control
(5.0 lpm)
Pressure
Regulator
CHARCOAL
SCRUBBER
Filter
PUMP
Flow Control
(1.0 LPM)
Pressure
Monitor

Figure 8-3: Pressure Calibration Monitor Points

07223B DCN6378