Teledyne API T703 Calibrator Operation Manual Getting Started
43

3.4.5. OPERATING MODES FOR THE O3 GENERATOR

The O3 generator can be set to operate in three different modes:

3.4.5.1. CNST (CONSTANT)

In this mode, the O3 output of the generator is based on a single, constant, drive voltage. There is no Feedback
loop control by the T703’s CPU in this mode.

3.4.5.2. REF (REFERENCE)

The O3 control loop will use the reference detector’s measurement of the O3 generator’s UV lamp as input. This
mode does not use the photometer to control the ozone generator.

3.4.5.3. BNCH (BENCH)

The O3 concentration control loop will use the photometer’s O3 measurement as input.
THIS IS THE DEFAULT AND MOST COMMON MODE OF OPERATION.
This setting will be the default mode of the T703 calibrator and will be mused whenever the calibrator is using
the GENERATE AUTO command or the GENERATE sequence step to create a calibration mixture. When
either the GENERATE MAN command or the MANUAL sequence step is active, the local O3 generator mode
(chosen during when the command/step is programmed) will take precedence.

3.4.6. SETTING THE O3 GENERATOR MODE

To select a default O3 generator mode, press: