Appendix C

Warranty/Repair Questionnaire

T703, M703E

(05747C DCN5798)
TELEDYNE API CUSTOMER SERVICE
Email: api-customerservice@teledyne.com
PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816
C-1
CUSTOMER:_______________________________ PHONE: _____________________________________
CONTACT NAME: __________________________ FAX NO. _____________________________________
SITE ADDRESS:____________________________________________________________________________
MODEL TYPE: ______________ SERIAL NO.:_________________ FIRMWARE REVISION:_____________
Are there any failure messages? _______________________________________________________________
_________________________________________________________________________________________
__________________________________________________________________________________________
__________________________________________________________________________________________
________________________________________________________________________ (Continue on back if necessary)
PLEASE COMPLETE THE FOLLOWING TABLE:

PARAMETER

Name in E-Series software
versions prior to v.C.1 Name in T-Series and
in E-Series w/software
v. C.1 and higher.

RECORDED VALUE ACCEPTABLE

VALUE

ACT ACT PPB 1% OF TARG
TARG TAR PPB 50 – 1000 PPB
OUTPUT FLOW OUT FLW LPM 2 – 5 LPM
REG PRESSURE REG PRES PSIG 15 ± 2 PSIG @ 5 LPM
BOX TEMP BOX TMP ºC 20 – 35 ºC
O3 GEN REF 1 O3GEN REF
1 mV 0 – 5000 mV
O3 GEN DRIVE 1 O3GENDRV
1 mV 0 – 5000mV
O3 LAMP TEMP 1 O3LAMPTMP1 ºC 48 ± 1 ºC
PHOTO MEASURE 2 PH MEAS
2 mV 2500 – 4700 mV
PHOTO REFERENCE 2 PH REF
2 mV 2500 – 4700 mV
PHOTO FLOW 2 PH FLW
2 LPM 0.720 – 0.880 LPM
PHOTO LAMP TEMP 2 PH LTEMP
2 ºC 58 ± 1 ºC
PHOTO SPRESS 2 PH SPRES
2 IN-HG-A -1” AMBIENT IN-HG-A
PHOTO STEMP 2 PH STEMP
2 ºC 25 – 48ºC
PHOTO SLOPE 2 PH SLOPE
2 1 ± 0.15
PHOTO OFFSET 2 PH OFFST
2 PPB 0 ± 10 PPB
Depending on options installed, not all test parameters shown will be avail able in your calibrator.
1 If ozone generator option inst alled.
2 If photometer option installed.
What is measured photometer flow rate ____________________________________________________ cc3/min
What is measured O3 generator flow rate? _________________________________________________ cc3/min
07223B DCN6378