Teledyne API T703 Calibrator Operation Manual T703 Calibration and Verification

137

8.2.2. VIEWING O3 GENERATOR CALIBRATION POINTS

To view these calibration points, press:

SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
Make sure that the
calibrator is in standby
mode.
SETUP X.X ENTER PASSWORD
818 ENTREXIT
Toggle to enter the correct
PASSWORD
DIAG SIGNAL I/O
PREV NEXT ENTR EXIT
DIAG O3 GEN CALIBRATION
PREV NEXT ENTR EXIT
Continue pressing NEXT until ...
DIAG O3 GEN CALIBRATION
CAL PNTS EXIT
Toggle to move between
calibration points
DIAG O3GEN 1) 500 MV, 5.0 MIN, ON
PREV NEXT INS DEL EDIT PRNT EXIT
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT

07223B DCN6378