General Troubleshooting & Repair Teledyne API T703 Calibrator Operation Manual

198

11.3.2.2. O3 Status LEDs

D15 (Green) - Photometer Lamp Heater

D16 (Yellow) – O3 Generator Lamp Heater

D2 (Yellow) Dry (zero) Air Pump Status

D9 (Green) External Zero Air Valve Status

D1 (RED)

Watchdog

Indicator

D7 (Green) Photometer Meas/Ref Valve Status

Figure 11-3: Relay PCA Status LEDS Used for Troubleshooting

Table 11-5: Relay PCA Status LED Failure Indications

SIGNAL I/O PARAMETER
LED FUNCTION ACTIVATED BY VIEW RESULT DIAGNOSTIC TECHNIQUE
D2
Yellow
Status of AC
powered Dry
Air Pump
ZERO_AIR_PUMP N/A
Pump should start /stop
Failed pump
Failed AC Relay on Relay PCA
Failed Relay PCA
Faulty AC Power Supply (PS2)
Faulty Connectors/Wiring
D7
Green
Photometer
Meas/Ref
Valve
PHOTO_REF_VALVE N/A
Valve should audibly change states.
If not:
Failed Valve
Failed Relay Drive IC on Relay PCA
Failed Relay PCA
Faulty +12 VDC Supply (PS2)
Faulty Connectors/Wiring
D9
Green
Status of DC
powered
Photometer
Pump
O3-PUMP-ON N/A
Pump should start /stop
Failed pump
Failed Drive IC on Relay PCA
Failed Relay PCA
Faulty AC Power Supply (PS2)
Faulty Connectors/Wiring
D15
Green
Photometer
Heater Status PHOTO_LAMP_HEATER PHOTO_LAMP_TEMP
D16
Green
O3 Generator
Heater Status O3_GEN_HEATER O3_GEN_TEMP
Voltage displayed should change.
If not:
Failed Heater
Faulty Temperature Sensor
Failed AC Relay
Faulty Connectors/Wiring

07223B DCN6378