Teledyne API - T100H Addendum to T100 Operation Manual Operating Instructions
31
6.2.4. Default DAS Channels
The default Data Channels included in the T100H analyzer’s software include the CONC, PNUMT & CALDAT
channels. The FAST & DETAIL preset channels are not included.
6.2.5. Remote Operation Using the External Digital I/O

6.2.5.1. Status Outputs

The function and pin assignment5s for the T100H digital status outputs are:.
Table 6-5: Status Output Signals
SATUS
CONNECTOR
PIN NUMBER1
STATUS
DEFINITION CONDITION
1 SYSTEM OK ON if no faults are present.
2 CONC VALID
OFF any time the HOLD OFF feature is active, such as during calibration or
when other faults exist possibly invalidating the current concentration
measurement (example: sample flow rate is outside of acceptable limits).
ON if concentration measurement is valid.
3 HIGH RANGE ON if unit is in high range of the AUTO Range Mode
4 ZERO CAL ON whenever the instrument’s ZERO point is being calibrated.
5 HIGH SPAN
CAL
ON whenever the instrument is set for DUAL or AUTO reporting range
mode an it’s high range span point is being calibrated .
6 DIAG MODE ON whenever the instrument is in DIAGNOSTIC mode
7 LOW SPAN CAL ON whenever the instrument is set for DUAL or AUTO reporting range
mode an it’s lows range span point is being calibrated .
8 SPARE
D EMITTER BUS The emitters of the transistors on pins 1-8 are bussed together.
SPARE
+ DC POWER + 5 VDC, 300 mA source (combined rating with Control Output, if used).
Digital Ground The ground level from the analyzer’s internal DC power supplies
1 Located on Rear Panel
07265A DCN6038