Calibration and Calibration Check Procedures Teledyne API - T100H Addendum to T100 Operation Manual)
34 STEP THREE: Perform the calibration according to the following flow chart:
Analyzer enters SPAN cal
mode and the SPAN key
appears.
The SPAN and CONC
buttons only appear when
the HIGH span gas is
selected in the previous
step
Press ENTR to changes
the OFFSET & SLOPE
values for the SO2
measurement.
Press EXIT to leave the
calibration unchanged and
return to the previous
menu.
ZERO CAL M STABIL= XXX.X PPM SO2=XXX.X
<TST TST> ZERO CONC EXIT
EXIT at this point
returns to the
SAMPLE menu.
Press ENTR to changes the
OFFSET & SLOPE values for
the SO2 measurement.
Press EXIT to leave the
calibration unchanged and
return to the previous menu.
Set the Display to show
the STABIL test function.
This function calculates
the stability o f the SO 2
measurement
Toggle TST> button until ...
Wait until STABIL falls
below 0.5 ppm.
This may take several
minutes.
ZERO CAL M STABIL= XXX.X PPM SO2=XXX.X
<TST TST> ENTR CONC EXIT
SPAN CAL M STABIL= XXX.X PPM SO2=X.XXX
<TST TST> SPANCO NC EXIT
SAMPLE SPAN TO CAL:HIGH
LOW HIGH ENTREXIT
Wait until STABIL falls
below 0.5 ppm.
This may take several
minutes.
SPAN CAL M STABIL= XXX.X PPM SO2=X.XXX
<TST TST> ENTR CONC EXIT
SPAN CAL M STABIL= XXX.X PPM SO2=X.XXX
<TST TST> ENTR CONC EXIT
SAMPLE RANGE = 5000.0 PPM SO2 =XXX.X
< TST TST > CAL CALZ CALS SETUP
Analyzers enters
ZERO cal
mode.
SAMPLE STABIL= XXX.X PPM SO2=XXX.X
< TST TST > CAL CALZ CALS SETUP
Allow zero gas to enter the zero port
at the rear of the analyzer.
SAMPLE STABIL= XXX.X PPM SO2=XXX.X
< TST TST > CAL CALZ CALS SETUP
Allow span gas to enter the high span port
at the rear of the analyzer.
SAMPLE STABIL= XXX.X PPM SO2=XXX.X
< TST TST > CAL CALZ CALS SETUP
07265A DCN6038