Teledyne API - T100H Addendum to T100 Operation Manual Calibration and Calibration Check Procedures

35

7.2. Manual Calibration Check with Ambient Zero and Two Ambient Span Valve Option (OPT 50C) Installed

Set up is identical to that shown in STEP ONE of the preceding section. To perform the zero/span check:

Record SO2 LOW span point reading
SAMP LE STABIL= XXX.X PPM SO2=XXX.X
< TST T ST > CA L SET UP
The ZERO, CONC and/or SPAN
buttons may appear at various
points of this process.
Do not press them this calibration
check procedure.
SAMPLE RANG E = 5000.0 PPM SO2 =XXX.X
< TST TST > CAL C ALZ CAL S SE TUP
Set the display to show
the STABIL test function.
This function calculates
the stab ility of th e SO 2
measurement
Toggle TST> button until ...
Wait u ntil STABIL falls
below 0.5 ppm.
This may take several
minutes.
Record SO2 zero point reading
SAMPLE SPAN TO CAL:HIGH
LOW HIGH ENTR EXIT
SAMPLE STABIL= XXX.X PPM SO2=XXX.X
< TST T ST > CAL CAL Z CALS SE TUP
Sets the analyzer’s
interna l valve s to
accept span gas
from the LOW
SPAN inlet.
SPAN CAL M STABIL= XXX.X PPM SO2=X.XXX
<TST TST> EXIT
SPAN CAL M STABIL= XXX.X PPM SO2=X.X XX
<TST TS T> EXIT
SAMPLE STABIL= XXX.X PPM SO2=XX X.X
< TST T ST > CAL CAL Z CALS SETUP
Record SO2 HIGH span point reading
SAMPLE SPAN TO CAL:HIGH
LOW HIGH ENTREXIT
SPAN CAL M STABIL= XXX.X PPM SO2=X.X XX
<TST TST> EXIT
Sets the analyzer’s
internal valves to
accept span gas
from the HIGH
SPAN inlet.
SAMPLE SPAN TO CAL:LOW
LOW HIGH ENTR EXIT
Input Zero gas
Flow Span Gas into Low Span Port
Flow Span Gas into Low Span Port