Topcom 200 HOW to USE, Display Error Messages, Time Settings, Talk Mode, Memory

Models: 200

1 88
Download 88 pages 6.89 Kb
Page 5
Image 5

TOPCOM EAR THERMOMETER 200

8HOW TO USE

• Please make sure to use a new and clean probe cover to ensure accurate measurement.

• Press the Start - button (1) to turn the power on. The display will shows all symbols.

Before inserting the probe:

- For children < 1 year : Pull the ear down and back.

- For children > 1 year to adult : Pull the ear up and back.

While holding the ear, insert the probe to seal the ear canal, then press the Start - button

(1)once. You will hear a short beep.

When you hear a short melody, the measurement is finished.Take out the thermometer

from the ear canal. The measured temperature is displayed. The result will be stored auto- matically into the memory. If the memory is full (30 records), the oldest measurement result will be removed.

It is recommended that you take three temperatures in the same ear under the following circumstances. If they are different, use the highest reading.

Infants in the first 3 months of life.

Children under three years old who have a condition such as acompromised

immune system and for whom the presence or absence of fever is critical.

When you are first learning to use the ear thermometer until you are comfortable with the technique and are gaining consistent readings

Important: As with any type of thermometer, slight temperature variations (+/- 0.2 - 0.3°C) can happen, if consecutive measurements are taken.

9DISPLAY ERROR MESSAGES

‘HI’ The measured temperature is higher than 42,9 °C.

‘LO’ The measured temperature lower than 32 °C

‘ErrE’ The thermometer temperature is not within the operation specifications (<10°C or >40°C)

‘ErrU’ Don’t measure after pressing ‘Measure’. Follow the steps mentioned in the user guide.

‘ErrH’ An error is detected during self-check. Remove the batteries for 10 seconds and try again.

’ Low battery. Replace the battery.

10TIME SETTINGS

Power on the unit by pressing the Start - button (1) once. You will hear a short beep. Time mode is displayed.

Press and hold the Memo - button (4). The hour digits are blinking.

Press the Memo - button (4) repeatedly to increase the digit.

Press the Talk - button (3) to confirm. The minutes digits are blinking.

Repeat previous steps for setting the minutes, year, month and day.

After setting the day, the unit will enter time mode.

11TALK MODE

The result of each measurement is broadcasted through the speaker.

To broadcast the time, press the Talk - button (3) in Time mode.

To disable the talk mode, press and hold the Talk - button (3) for 2 seconds. You hear one beep tone.

To enable the talk mode, press and hold the Talk - button (3) again for 2 seconds. You hear a double beep tone.

12 MEMORY

The Ear Thermometer will automatically store the last 30 measurements.

Power on the unit by press the Start - button (1) once. You will hear a short beep. Time mode is displayed.

Press the Memo - button (4). The last result is displayed.

Press the Memo - button (4) repeatly, the display will show the temperature of the previous measurements.

Press and hold the Memo - button (4) repeatly to scroll rapidly.

ENGLISH

5

Page 5
Image 5
Topcom 200 manual do utilizador HOW to USE, Display Error Messages, Time Settings, Talk Mode, Memory