15 DETACH/ATTACH OF TRIBRACH
15-1
15 DETACH/ATTACH OF TRIBRACHThe instrument is easily detached or attached to the tribrach, with a tribrach locking lever loosened or
tightened for this purpose.
●Detachment
1Loosen the tribrach locking lever, by revolving it 180° or 200gon in the counterclockwise direction
(which will point the triangle mark upwards).
2Grip the carrying handle firmly with one hand while holding the tribrach with the other. Then lift the
instrument straight upwards and off.
●Attachment
1Hold the instrument by the carrying handle, with one hand, and carefully lower it on top of the
tribrach while, at the same time, coinciding the alignment piece with the tribrach alignment groove
on the instrument and tribrach respectively.
2When fully seated, revolve the tribrach locking lever 180° or 200gon clockwise ( which will point the
triangle mark downwards again).
3Confirm that the upper instrument section is securely attached.
●Locking the Tribrach Locking Lever
The tribrach locking lever can be locked, to prevent it be accidentally removed, especially if the
upper instrument section is not being detached very often. Simply tighten the securing screw on the
locking lever with the accessory screwdriver, found in the case.
Alignment piece
Securing screw Tribrach locking lever
Alignment groove