6 SPECIAL MODE (Menu Mode)
6-18
6.4 Setting Mode 1In this mode, the following settings are possible.
1. Setting Minimum Reading
2. Auto Power off
3. Vertical and Horizontal Angle Tilt Correction (Tilt ON/OFF)
(GPT-2009 has vertical angle tilt correction only.).
4. Systematic Error of Instrument Correction (GPT-2003/2005/2006)
5. Selecting battery type
●This setting is memorized after power off.
6.4.1 Setting Minimum Reading
Select minimum display unit for angle measurement, coarse distance measurement mode.
●To select minimum display for fine measurement mode, see Chapter 16 “SELECTING MODE” .
[Example] Minimum angle : 5”, Coarse : 1mm
Models Angle Unit Coarse mode
Distance unit
Degree GON MIL
GPT-2003
GPT-2005
GPT-2006 5" / 1" 1mgon / 0.2mgon 0.1mil / 0.01mil 10mm (0.02ft)/
1mm(0.005ft)
GPT-2009 10" / 5" 2mgon / 1mgon 0.1mil / 0.01mil
Operating procedure Operation Display
1After pressing the [MENU] key, press the [F4](P
↓
)
key twice to get the menu on page 3. [MENU]
[F4]
[F4]
2Press the [F1] key. [F1]
3Press the [F1] key. [F1]
4Press the [F1] key. [F1]
5Press the [F2](5") key and press the [F4](ENTER)
key. [F2]
[F4]
6Press the [F2] key. [F2]
MENU 3/3
F1:PARAMETERS 1
F2:CONTRAST ADJ.
P
↓
PARAMETERS 1
F1:MINIMUM READING
F2:AUTO POWER OFF
F3:TILT P
↓
MINIMUM READING
F1:ANGLE
F2:COARSE
MINIMUM ANGLE
[F1: 1" ]
F2: 5"
ENTER
MINIMUM READING
F1:ANGLE
F2:COARSE
COARSE READING
F1: 1mm
[F2:10mm]
ENTER