TIC-LF654L
15
Table 5. Specification Code for converters Model Specification Code Contents LF620
type
LF622
type
1 2 3 4 5 6 7 8 9 1011 12 13 14
L F 6 2 Electromagnetic flowmeter converter
0 Combined (Integral) type
–
2 Separate (Remote) type
–
F
Purpose
cFMus class I, Division 2 approved
A
Shape
Standard type with case
A
C
E
Converter mounting fitting
None
Panel, Accessory for wall mounting (BNP material: SUS304)
Accessory for pipe installation (BNP material: SUS304)
–
–
2
Digital input/output
Digital output points 2 (DO1+DO2) +Digital input point 1 (DI)
1
2
3
Current output and Communication function(Note1)
Current output + HART communication
PROFIBUS communication
Current output + Modbus (RS485) communication
1
2
3
Power supply(Note2)
100Vac-240Vac 50/60Hz, 110Vdc
24Vdc
110Vdc
E
Instruction manual
English
Code explanation: : Standard : Option
–
: Not available
Note 1: When PROFIBUS communication is provided, current output(4-20mA) and HART communication cannot be used.
When Modbus communication is provided, digital output points 1(DO1) and digital output points (DO2),
digital input point 1(DI), HART communication cannot be used.
Refer to Table 6 for more details.
Note2: Select 110Vdc for test report inspected under the condition of 110Vdc.
Table6. Communication functions and output selection table
Selection of Function Availability of outputs
Code
(10th digit)
Selected
Communication 4-20mAdc DO1 DO2 DI
1 HART
2 PROFIBUS
3 Modbus
(Note)