Toshiba LQ500 specifications Converter Specifications

Models: LQ500

1 9
Download 9 pages 8.7 Kb
Page 3
Image 3

LQ500

200 mm (8”)

8 mS/cm maximum

250 mm (10”)

8 mS/cm maximum

300 mm (12”)

6 mS/cm maximum

Note 1: The LQ500 can not havean accurate density (consistency) measurement when it is over the specification according to reduce the microwave signal.

Note 2: The LQ500 density (consistency) measurement for application where liquids containing highly conductive particles such as active carbon and metal particles may be affected. Consult Toshiba for detail when the measuring liquid contains such particles.

Wetting materials:

Name

Materials (*1)

Main pipe

SCS14A cast (equivalent to

316 SS) (standard) (*2, *3)

 

Temperature detector sheath

316 stainless steel (*4)

Applicator window

Polysulfone (*4)

Applicator window sealant

Fluoric rubber

*1 Avoid using the LQ500 for applications where harmful liquids that cause corrosion, deterioration, or changes in quantity for the wetting materials are used. Make sure all materials at these wetting parts that are suitable for your CIP or not before cleaning.

*2 The smoothness inside the pipe on this material is;

Type

Meter size

Smoothness

 

50 to 200 mm

No buffing

Standard type

(1/2” to 8”)

 

250 & 300 mm

Buffing # 150

 

 

(10” & 12”)

 

 

Option type

50 to 300 mm

Buffing # 150

(1/2” to 12”)

 

 

*3 State the wetted materials when you choose these options.

*4 The materials of them are changed for abrasive applications.

Need to choose the specification code for this application.

Applicator:

Serves as an antenna to send and receive microwave signals, one set provided.

Temperature detector: RTD (Pt100)

Fitting:

Direct fitting to vertical or horizontal piping. (Refer to the section of Piping Precautions.)

Converter Specifications

Output signals

Density (consistency) measurement output:

4-20mAdc (load resistance 750 ohm maximum, isolated output.)

Density (consistency) fault or Maintenance signal: 125Vac, 0.1A (resistive load) solid state contact; opens when an error occurs in the converter or when the LQ500 is in the setting change mode, otherwise the contact remains closed.

Communication signal:

Digital signal is superimposed on 4-20mAdc

current signal (conforming to HART protocol(*1)). Load resistance:240 to 750 ohm

Load capacity:0.25μF maximum

*1 HART (Highway Addressable Remote Transducer) protocol is a communications protocol for industrial sensors recommended by HCF (HART Communication Foundation).

Note: The optional AF900 hand-held terminal can be used to operate the LQ500 from remote places by connecting the AF900’s probe lead between the LQ500’s 4-20 mA dc output signal lines.

Input signals

External synchronized input signal: In order to avoid problems of density (consistency) measurement such as inhomogeneous condition caused by discontinuous process operation and empty pipe condition caused by stopping process operation. <Specification>

One dry “make” contact;

Contact capacity of 24Vdc with 0.1 to 2.0A is required.

This contact signal can be used to start or stop density (consistency) measurement in synchronization with an external contact, such as the contacts on a pump.

The measurement starts or stops as follows: Contact closed: Starts density (consistency) measurement.

Contact open: Stops density (consistency) measurement.

Density multiplier switching signal:

In order to achieve selecting up to 4 kinds of liquid concentration measurement independently as maximum.

<Specification>

Two voltage signals described below are required: Input voltage: H level: 20 to 30 Vdc

L level: 2 Vdc or less

Input resistance: Approx. 3k ohm

Conductivity correction signal:

Need to prepare an additional conductivity meter when using this function. Install in where is able to have a stable and accurate measurement.

<Specification>

Input signal: 4 to 20mAdc

Conductivity range: 0 to 10mS/cm

Update period for density (consistency) measurement output and display:

Approx. 1 second

Functions by software as standard:

Data saving function:

In order to save measurement data into the memory of converter temporary.

The oldest data is overwritten. <Specification>

Data storage points: 256 points maximum.

Period: 1 to 1,800 minutes (1 minute each). ex 1: The data is saved for approx. latest 4.26

hours when programming every minute. ex 2: The data is saved for approx. latest 21.3

3

Page 3
Image 3
Toshiba LQ500 specifications Converter Specifications