Toshiba LQ500 specifications Detector Specifications

Models: LQ500

1 9
Download 9 pages 8.7 Kb
Page 2
Image 2

LQ500

range, which is the upper limit setting range.

Resolution:

Meter size

50 mm (2”)

80 mm (3”)to

300 mm (12”)

 

 

Resolution

0.002%TS

0.001%TS

Note 1: Above values are the results of commuting ability in the phase measurements of the converter.

Note 2: Density (consistency) determination resolution for sample reagent;

Meter size

50 mm

80 mm (3”)to

(2”)

300 mm (12”)

 

Density(consistency)

0.1%TS

0.05%TS

determination repeatability

*The density (consistency) determination resolution stated above is defined due to manufacturing limitation to make reagents with stable distribution and a minimum difference of fluid density (consistency).

<Notice>

1.Install a sample tap near the LQ500 as close as possible to get an accurate density (consistency) measurement using the LQ500.

2.Take several samples in rapid sequence for more accurate density (consistency) measurement with less human error factors.

Environmental conditions:

Items

Temperature range

Humidity range

 

Standard: 0 to 50 deg.C

 

Detector

(32 to 122 deg.F)

5 to 85%RH

Option: -20 to 50 deg.C

 

 

(4 to 122 deg.F) (*1)

(no condensation)

Converter

0 to 50 deg.C

 

( 32 to 122 deg.F)

 

 

 

*1 A rubber heater is installed at the RF part of the detector.

Need to prepare an additional power supply either 100Vac type or 200Vac type for it by yourself. Determine one of them at the specification code. This option is available for detector only.

Structure: Converter;IP65,

Detector:IP67, Watertight

Note: Outdoor installation is possible. However, provide a sunshade for the converter section if direct sunlight is unavoidable.

Microwave power: 10 mW

Vibration resistance:

No resonance to the following levels of vibration:

(1)No failure for 5 to 150 Hz with the following acceleration in each device for 30 minutes in each axis of X, Y, and Z (90 minutes); Converter: 4.9 m/s2, Detector: 25m/s2

(2)No failure for 5 to 150Hz with the following acceleration in each device for 3 minutes in each axis of X, Y, and Z, 10 times (90 minutes as total);

Converter: 4.9 m/s2, Detector: 25m/s2

Note: Avoid using the LQ500 in an environment with constant vibration.

Cables: See the Table 2

Table 2. Cable specifications

 

Cables Between detector

Power

 

(RF part)and converter

Specifications

supply

(*1)

cable of

 

Power supply

Communi-

LQ500 (*2)

 

(24Vdc)

cation

 

Cable type

CVVS-2C-2S

CVVS-5C-

CVV-3C-2S

 

 

1.25S

 

Cross-sectional

2 mm2

1.25 mm2

2 mm2

area (*3)

 

 

 

Number of cores

2

5

3

Cable diameter (*4)

11 to 13

11 to 13

11 to 13

mm2

mm2

mm2

*1 10m (32.8 ft) length is packed as standard.

*2 Need to prepare this cable by the customer.

*3 Need to use a sheathed cable.

*4 If the diameter of the cable is smaller than the inside diameter of the packing, enlarge the cable diameter to the same size as the packing by wrapping around the cable.

This dimension is coming from a diameter of the cable gland of the LQ500.

Conformance to European Community Directives:

EMC directive 89/336/EEC

The low voltage 93/68/EEC

PED 97/23/EC (Note l)

Note: See table 6 in detail.

Approved hazardous locations certification:

UL/CUL Class I, Division 2, Groups A, B, C and D (UL/CULonly for explosion proof)

Weight: Refer to Outline Dimensions (Table 3).

Part 15 of the FCC rules: Certified.

Detector Specifications

Meter size: 50mm (2”), 80mm (3”), 100mm (4”), 150mm (6”), 200mm (8”), 250mm (10”), and 300mm (12”)

Flange standard and maximum working pressure:

Flange standard

Maximum working pressure

ANSI Class 150

1 MPa (150 psi)

DIN10 and BS10

1 MPa (10 bar)

DIN 16

1.6 MPa (16 bar)

JIS 10K

1 MPa (10 kgf/cm2)

Note: Each product was passed a hydraulic test under twice pressure rate for 15 minutes toward the specification.

Fluid temperature:

0 to 100 deg.C (32 to 212 deg.F) without freezing and bubbles conditions

Allowable fluid conductivity:

Meter size

Fluid conductivity

50 mm (2”)

20 mS/cm maximum

80 mm (3”)

16 mS/cm maximum

100 mm (4”)

15 mS/cm maximum

150 mm (6”)

10 mS/cm maximum

2

Page 2
Image 2
Toshiba LQ500 specifications Detector Specifications