VXI CT-400 user manual Connecting the Mainframe to Earth Ground, AIR Flow Requirements

Models: CT-400

1 57
Download 57 pages 24.99 Kb
Page 26
Image 26

VXI Technology, Inc.

CONNECTING THE MAINFRAME TO EARTH GROUND

The CT-400 is configured at the factory for 115 V ac, 60 Hz. The supplied power cord grounding conductor provides adequate grounding for this voltage and frequency level. However, if the CT-400 is to run at anything higher than 63 Hz, the mainframe MUST be connected to earth ground at the mainframe as follows:

1)Connect a 16 AWG (or larger) wire to the ground terminal located on the rear panel (this connection is marked by a symbol). Use a grounding lug with a star washer or a toothed lug, to make this connection. See Figure 2-1.

2)Attach the other end of the wire to a permanent earth ground also using a star washer or a toothed lug.

To avoid electrical shock, the mainframe ground terminal MUST be connected to earth ground for frequencies higher than 63 Hz.

Chassis Ground Terminal

Fuse

-DO NOT REMOVE THIS POWER MODULE

!THIS INSTRUMENT HAS AC POWER APPLIED

-THE GROUNDING CONNECTION IN THE

!CORD MUST BE CONNECTED TO ENSURE PROTECTION FROM ELECTRICAL SHOCK.

-TO AVOID ELECTRICAL SHOCK, DISCONNECT

!AC POWER CORD PRIOR TO REPLACING FUSE.

- FOR CONTINUED FIRE PROTECTION ONLY WITH A FUSE OF THE CORRECT AND RATING.

F200

 

 

15A T 250 (VAC)

REMOVE COVER TO

 

 

SERVICE AIR FILTER

 

J200

PWS

TEMP

100 - 240 VAC

AMB

TEMP

15 A MAX.

PWS

PWR

47 -63 Hz

PWS

VOLT

 

PWS

CUR

 

FAN

SPEED

FAN SPEED

HIGH

VAR

LOW

J201

MONITOR / CONTROL

J211

J210

FIGURE 2-1: FUSE AND GROUND LOCATION

AIR FLOW REQUIREMENTS

The CT-400 must be positioned to provide adequate airflow through the unit. Whether used as a bench-top or rack mount unit, provide at least a one inch clearance, at the back and sides, to allow for adequate airflow. Two air filters are installed in the mainframe at the factory, one for the power supply fan (right side from rear view) and a larger one for the mainframe fans (in the rear at the bottom). These filters may be removed if additional airflow is required, but only if the air is filtered at another source within the rack or if it is installed in a clean environment.

26

CT-400 Installation

Page 26
Image 26
VXI CT-400 user manual Connecting the Mainframe to Earth Ground, AIR Flow Requirements

CT-400 specifications

The VXI CT-400 is an advanced test and measurement platform designed to cater to the needs of engineers and technicians across various industries, including telecommunications, aerospace, and defense. Known for its robust performance and versatility, the CT-400 offers a comprehensive range of features that streamline testing processes, ensuring accuracy and efficiency.

One of the main features of the VXI CT-400 is its modular architecture. This design allows users to configure the system according to their specific requirements, accommodating different testing applications. With a variety of interchangeable modules, users can easily customize their setup for a wide range of tests, from RF and microwave measurements to digital signal processing.

The CT-400 leverages cutting-edge technologies to deliver precise results. Equipped with high-resolution analog-to-digital converters (ADCs) and digital-to-analog converters (DACs), it ensures fidelity in the measurement and reproduction of signals. The device also integrates advanced signal processing algorithms that facilitate enhanced data analysis, making it easier to interpret complex measurement results.

Another notable characteristic of the CT-400 is its support for various communication protocols. This flexibility allows engineers to seamlessly integrate the device into existing test environments, making it compatible with other testing equipment and software platforms. Whether dealing with legacy systems or modern applications, the CT-400's interoperability plays a crucial role in enhancing productivity.

In terms of user experience, the VXI CT-400 is designed with an intuitive interface that simplifies operation. A large, high-resolution display provides clear visualization of measurements, while easy-to-navigate menus enable swift access to different functions. This user-centric design minimizes the learning curve, making it accessible for both experienced professionals and newcomers in the field.

Furthermore, the CT-400 is built with durability in mind. Its rugged construction ensures reliable operation in diverse environments, including harsh industrial settings. This resilience is complemented by robust software support, enabling regular updates and the addition of new features to keep the device up-to-date with technological advancements.

In conclusion, the VXI CT-400 stands out as a powerful test and measurement solution characterized by its modular design, advanced technologies, versatile communication capabilities, intuitive interface, and rugged construction. With these attributes, it serves as an indispensable tool for engineers striving for precision and efficiency in their testing processes.