7.8.3 Digital Format

The Digital format was implemented as an easy way to design purely digital waveforms, i.e. waveforms that are either high or low. The digital format provides a very efficient way of representing waveforms that assume only a high or low value.

If the value of the data point is 0, then the SIG Out output is set to its minimum negative output voltage and the SYNC Out output is set to the logic Low state (0V) for that point.

If the value of the data point is non-zero, then the output is set to its maximum positive output voltage and the SYNC Out output is set to the logic High state (+5V) for that point.

All rules for this format are identical to those for Floating Point Format, except that the Time or Point Number value need not be between +1.0 and -1.0 in value since the 4071 only sees if the value is zero or nonzero. Also, the “P” character has no meaning since the SYNC Out output is controlled by the data itself.

With this format, it is possible to specify each point of a digital waveform with only two bytes (a “1” or a “0” and a data separator character). This makes this format as efficient as binary format for data downloads. However since the data consists entirely of ASCII characters, it is easier to work with than binary format.

Example

This twelve point digital waveform

0 1 0 0 1 1 0 0 0 1 1 1

will render the following waveform on both the SIG Out and SYNC Out connectors:

56

BK Precision 4071 User Manual Rev.2.2

Image 61
Contents User’s Manual Page +K Precision Corp Model User’s ManualTable of Contents Remote operation Warranty informationArbitrary waveform system DC Operation OptionIntroduction Description Modes Feature summarySync out connector Installation and setupSIG Out connector External Modulation In connector External Trigger/Gating/FSK/BPSK input External Arb Sync In / Out connector External Reference Input / Ext Arb. Clock connectorRS232 Interface connector Line in connectorOperating controls and keys Mode keyField arrow keys Recall / Store keyNumeric keys 0 to 9, Trigger keyÉ, ê, è, ç keys Clear key One Touch Mode KeysMHz/dBm, KHz/Vp-p/Sec, Hz/mVp-p/mS keys Operating guide Operating mode selectionChanging frequency Changing levelModifying an existing value Changing valuesEntering a new value Basic Sinewave CW Mode Mode descriptionsSinewave Mode 000,000.00 Hz 10.0 dBm Internal AM Mode External AM Mode Parameters External AM ModeExternal AM Input gain 000,000.00 Hz PEP level -10.0 dBm Input GainInternal FM Mode Parameters Internal FM ModeInt FM 1,000 Hz Pk dev 10,000 Hz 000,000.00 Hz 10.0 dBm Peak Frequency DeviationExternal FM Mode Parameters External FM ModeExternal FM Pk dev 10,000 Hz 000,000.00 Hz 10.0 dBm Internal PM Mode Parameters Internal PM ModePeak Phase Deviation Int PM 1,000 Hz Pk dev 180 deg 000,000.00 Hz 10.0 dBmExternal PM Mode Parameters External PM ModeExternal PM Pk dev 180 deg 000,000.00 Hz 10.0 dBm Sweep Mode Sweep time Up / Down sweepInternal FSK Mode Parameters Internal FSK ModeMark Frequency Space FrequencyExternal FSK Mode Parameters External FSK ModeExt FSK Mark 1,000,000.00 Hz Space 2,000,000.00 Hz 10.0 dBm Ext FSK InputBurst Mode Trig Ton 10,000 mS 000,000.00 Hz Off 10.0 dBm Burst Mode IntroductionBurst frequency Internal SSB Mode Parameters Internal SSB ModeInt SSB 1,000 Hz Upper Sideband 000,000.00 Hz 10.0 dBm Upper/Lower Sideband selectionsExternal SSB Mode Parameters External SSB ModeExternal SSB Upper Sideband 000,000.00 Hz 10.0 dBm Dtmf generation # Duration 200 mS Delay Level -10.0 dBm Dtmf Generation ModeDtmf Generation Mode Parameters BackgroundDtmf Duration Dtmf DelayDtmf Detection Mode Parameters Dtmf Detection ModeDtmf detection on Ext Mod input 3 # a Dtmf detected digitsPower & Voltage Measurement Mode Parameters Power & Voltage Measurement ModePower Level Voltage LevelRemote Mode Arbitrary ModeOther mode Internal Bpsk Mode Parameters Internal Bpsk ModeInt Bpsk 1,000 Hz 000,000.00 Hz 10.0 dBm External Bpsk Mode Parameters External Bpsk ModeExternal Bpsk 000,000.00 Hz 10.0 dBm 10,000 Hz F2 Ton 000 mS Toff 10.0 Dualtone Generation ModeDualtone Examples Using the Data Modulation Mode Data Modulation ModeData Modulation Mode Parameters Baud RateExample Message Downloading Modulation Message DataSaving the Message Triggering the TransmissionVCO Mode Parameters Voltage Controlled Oscillator VCO ModeStart End 1,000,000.00 Hz 10.0 dBm End FrequencyRemote Operation ConnectingChecking your connection with Hyper Terminal Remote Control Programming Rules Remote Control OperationReset 4071 to Sine wave mode Remote Control Commands× Ø Returns CE1,0 Enable, Disable LCD echo to terminal K1.0 Enable, Disable front panel keys and rotary knobReport hardware and software versions Linear Trig Up Time 10,000 mS Stp 31,500,000.00 HzF0-9 Move cursor to field 0 to ? or H Print help menuExample Remote Control ExamplesQuick Start Guide Arbitrary Waveform ModePin 9-pin cable Pin 25 pin cable adapter Description of the Arbitrary Waveform Generator Introduction to the Arbitrary Waveform SystemFeature Summary Switching ModesFunction Generator Pulse GeneratorInt Clock / Ext Clock / Lock Master Lock slave mode Arbitrary Waveform ModePhase Continuous/Triggered modeFunction Gen Wave Pos Ramp Rep Freq 100,000 Hz 000 mV Function Generator ModeWaveform Repetition FrequencyPositive Only Pulse Generator ModeDuty Cycle Downloading arbitrary waveforms Using WAVELOAD.EXESaving Arbitrary Waveforms Cwaveload /COMx /y /baudz filename.extUsing your own program Floating Point Format Data FormatsRules for Floating Point format Example #2 Time & Value Floating Point FormatDigital Format Rules for Integer format Integer FormatRules for Hexadecimal format Hexadecimal FormatExample The data value E468 is handled as follows Rules for Binary format Binary FormatExample Arb Multiple Units LockingGenerator Phase Resolution More on Phase OffsetsStarting Address = Number of waveform points * Phase Offset Waveform Size RestrictionsProgram Listing ARB.BAS Example Arb ProgramClose System Subroutines Return Print #1, a Send integer value to Return Return DC Operation Option Wiring DiagramOperating Modes 4071 SpecificationsRemote Software CDAppendix a Hardware SetupCLS CLS Print Print Subroutines END SUB Appendix B Television Remote Control ExampleCLS Make sure a 4071 is attached to the serial port Print Limited Two-Year Warranty Service Information Savi Ranch Pkwy Yorba Linda, CA 92887 USA TEL