Meade LXD55 instruction manual D e f g

Page 41

Tip:

A useful tool for making collimation easier is to make a hole in the plastic cap that comes with the eyepiece to protect it during shipping. Use a drill to make a 1/8" hole in the center of the cap and place it over the eye- piece holder for use during the collimation procedure.

bFocuser drawtube

cDiagonal mirror

dReflection of primary mirror

eReflection of secondary mirror (darkened due to back lighting)

fPrimary mirror clips

gReflection of observer's eye

Fig. 31b: The four collimation screws on the diagonal mirror housing.

knurled knob

thumb screw

Fig. 31c: The six collimation screws on the rear of the primary mirror cell.

b c d e f g

Fig. 31a: The view you will see while collimating a Schmidt-Newtonian telescope.

shipment of your telescope. It is only necessary for you to confirm that the telescope has not been badly jarred out of collimation, and to perform the final fine-tuning of Step 4, below.

Fig. 31a shows a correctly collimated Schmidt-Newtonian telescope, as it appears when viewed through the focuser with the eyepiece removed.

To check and, if necessary, set the optical collimation, follow these steps.

1.Observe through the focuser and orient your body so that the telescope's primary mirror is to your right, and the correcting plate end of the telescope tube is to your left. The diagonal mirror will appear centered as shown (2, Fig. 31a). If the diagonal appears off center, then adjust the 4 collimation screws on the plastic diagonal mirror housing.

2.If the reflection of the primary mirror (3, Fig. 31a) is not centered on the surface of the diagonal mirror, adjust the 4 collimation screws on the plastic diagonal mirror housing to center the reflection.

As described above, the 4 collimation screws (Fig. 31b) on the plastic diagonal mirror housing are used for two different adjustments during the collimation procedure.

Important Note: Do not force the 4 screws past their normal travel, and do not rotate any screw or screws more than 2 full turns in a counter- clockwise direction (i.e., not more than 2 full turns in their "loosening" direction), or else the diagonal mirror may become loosened from its support. Note that the diagonal mirror collimation adjustments are very sensitive: generally turning a collimation screw 1/2-turn will have a dra- matic effect on collimation.

3.If the reflection of the diagonal mirror is not centered within the reflection of the pri- mary mirror, adjust the 3 collimation screws located on the rear of the primary mirror cell.

Note: There are 6 screws (Fig. 31c) on the primary mirror cell. The 3 knurled knobs are the collimation screws, and the 3 smaller thumb screws are locking screws. The locking screws must be loosened slightly in order to adjust the collimation screws.

Proceed by "trial and error" until you develop a feel for which collimation screw to turn in order to change the image in any given way.

41

Image 41
Contents Meade Instruments Corporation Page Contents 1 1% g f e Telescope Features$2% LXD55-SeriesYour Personal Window to the Universe Computer Control Panel see e LXD55 Tips Tour the Cosmos with Just the Push of a Button Autostar FeaturesFeatures Tip Telescope Assembly How to Assemble Your TelescopeAccessories Getting StartedViewfinder assembly Schmidt-Newtonian models Insert eyepiece into holder and tighten thumbscrews Balancing the TelescopeChoosing an Eyepiece Aligning the ViewfinderObserving by Moving the Telescope Manually Too Much Power?Slew Speeds Autostar Models Only Activate the Arrow Keys Autostar Models OnlyObserve the Moon Setting the Polar Home PositionTracking Objects 16bUsing Autostars GO to Capabilities Autostar Models Only Observe a Star using the Automatic Tracking FeatureInitializing Autostar Autostar Models Only Moving Through Autostar’s Menus Autostar Models OnlyTraining the Drive Autostar Models Only Go To Saturn Autostar Models Only Easy Alignment Autostar Models OnlyPlanet Saturn is over 800 million miles from the Earth Using the Guided Tour Autostar Models OnlyTo Calculate Sunset time Autostar Navigation ExerciseNavigating Autostar Entering Data into AutostarSelect Item Object Autostars MenusObject Menu MenusGlossary Menu Event MenuUtilities Menu LXD55-SeriesEnter Setup MenuObserving Considerations Setup Menu Daylight SAVING, Surf the Web To Edit a Site Adding Observing SitesTo Add a Site to the user-defined site list To GO to a user-entered object Using Autostar to Find Objects Not in the LibrariesTo Add a landmark to the database Observing SatellitesLandmarks To perform a Landmark Survey To Check on the Available MemoryTo Select a landmark from the database Few tips on photography with the LXD55-Seriestelescopes Photography10 f/4 Optional AccessoriesSeries 4000 Photo-Visual Page General Maintenance MaintenanceD e f g Page Defocused star images. Misaligned 1, 2, Aligned Inspecting the Optics Meade Customer Service4 Model SN-8, SN-8EC Specifications Specifications5 Model SN-6, SN-6EC Specifications 8 Model AR-6 Specifications 10 f/4 Model SN-10 Specifications9.3 Model AR-5, AR-5EC Specifications Autostar Dimensions 10 Model SC-8 SpecificationsAutostar Specifications Appendix a Celestial Coordinates Setting CirclesOne- and Two-Star Polar Alignment Locating the Celestial PoleOne-Star Polar Alignment Two-Star Polar AlignmentProcedure #1 To correct for attitude misalignment Axis Alignment ProcedurePolar Alignment Viewfinder Method 1 Manual AlignmentPosition a Procedure #2 To correct for elevation misalignmentTips Star ChartsMethod 2 Axis Alignment Using Autostar Latitude Chart for Major Cities of the World Appendix C Latitude ChartCity State/Prov./Country Latitude City Country LatitudeAppendix D EC Handbox Slew Speeds To move the telescope using the Electronic ControllerModes of Operation Tracking Rates Optional Autostar Computer Controller for LXD55 EC Users Changing the Polar ModeMounting the SC bracket and optical tube to the mount How to Attach the Optical Tube to the MountAutostar Glossary Appendix F Basic AstronomyObjects in Space MoonDeep-Sky Objects PlanetsFurther Study Page Meade Instruments Corporation