1. Important
Guidance and manufacturer’s declaration – electromagnetic immunity – for all EQUIPMENT and
SYTEMS:
The monitor is intended for use in the electromagnetic environment specified below.The customer or the user of the monitor should assure that it is used in such an environment.
Immunity test | IEC 60601 test level | Compliance level | Electromagnetic environment – | |
guidance | ||||
|
|
| ||
Electrostatic | 6 kV contact | 6 kV contact | Floors should be wood, concrete or | |
discharge (ESD) | 8 kV air | 8 kV air | ceramic tile. If floors are covered with | |
IEC | synthetic material, the relative humidity | |||
|
| |||
|
| should be at least 30%. | ||
|
|
| ||
|
|
|
| |
Electrical fast | 2 kV for power supply | 2 kV for power supply | Mains power quality should be that | |
transient/burst | lines | lines | of a typical commercial or hospital | |
IEC | 1 kV for input/output | 1 kV for input/output | environment. | |
| ||||
| lines | lines |
| |
|
|
|
| |
Surge | 1 kV line(s) to line(s) | 1 kV line(s) to line(s) | Mains power quality should be that | |
IEC | 2 kV line(s) to earth | 2 kV line(s) to earth | of a typical commercial or hospital | |
environment. | ||||
|
|
| ||
interruptions and | <5 % UT (>95 % dip in | <5 % UT (>95 % dip in | Mains power quality should be that | |
voltage variations | UT) for 0.5 cycle | UT) for 0.5 cycle | of a typical commercial or hospital | |
on power supply | 40 % UT (60 % dip in | 40 % UT (60 % dip in | environment. If the user of the monitor | |
input lines | requires continued operation during | |||
UT) for 5 cycles | UT) for 5 cycles | |||
IEC | power mains interruptions, it is | |||
|
| |||
70 % UT (30 % dip in | 70 % UT (30 % dip in | recommended that the monitor be | ||
| ||||
| UT) for 25 cycles | UT) for 25 cycles | powered from an uninterruptible power | |
| <5 % UT (>95 % dip in | <5 % UT (>95 % dip in | supply or a battery. | |
|
| |||
| UT) for 5 sec | UT) for 5 sec |
| |
|
|
|
| |
Power frequency | 3 A/m | 3 A/m | Power frequency magnetic fields should | |
(50/60 Hz) |
|
| be at levels characteristic of a typical | |
magnetic field |
|
| location in a typical commercial or | |
IEC |
|
| hospital environment. | |
|
|
| ||
|
|
|
|
Note
UT is the a.c. mains voltage prior to application of the test level.
3