Specifications
Tripp Lite has a policy of continuous improvement. Specifications are subject to change without notice.
SmartPro® & OmniSmart™
Medical-Grade UPS Systems
Model: | OMNISMART350HG | SMART700HG | SMART1200XLHG |
Series: | AGOM350KSR57HG | AGSM700PSR3HG | AGSM1200PSR3HG |
Output Capacity (VA/Watts): | 330/225 | 700/450 | 1000/750 |
Battery Runtime (Half Load/Full Load) Minutes: | 46/18 | 42/18 | 34/16+ |
Battery Recharge Time: | 2-4 hrs. | 2-4 hrs. | 2-4 hrs. |
Approvals: | UL2601-1, UL1778 | UL2601-1, UL1778 | UL2601-1, UL1778 |
| FCC Class B and FCC Part 68 | FCC Class A | FCC Class A |
Tel/DSL Protection: | 2 lines | — | — |
| | | |
OmniSmart™ | | | |
UPS Systems with Hospital-Grade Plug & Receptacles | | |
Model: | OMNISMART450HG | OMNISMART700HG | OMNISMART1400HG |
Series: | AGOM700KSR61 | AGOM700KSR61 | 163655U |
Output Capacity (VA/Watts): | 450/280 | 700/425 | 1400/940 |
Battery Runtime (Half Load/Full Load) Minutes: | 17/5 | 22/9 | 24/8 |
Battery Recharge Time: | 2-4 hrs. | 2-4 hrs. | 2-4 hrs. |
Approvals: | UL1778 | UL1778 | UL1778 |
| FCC Class B | FCC Class B | FCC Class B |
| | | |
ALL MODELS: Input Voltage/Frequency (120V/60 Hz); Output Waveform Line Mode (filtered sine wave); Output Waveform Battery Mode (PWM sine wave); AC Surge Suppression (exceeds IEEE 587 Cat. A & B standards); AC Noise Attenuation (>40 dB); AC Protection Modes (H to N, H to G, N to G).
+Battery runtime can be extended with addition of optional Tripp Lite External Battery Packs (model BP36V27, sold separately). External batteries will increase both the battery runtime and the battery recharge time.
UL CLASSIFICATION FOR MEDICAL EQUIPMENT:Class I, Type B, Ordinary Equipment, Continuous Operation.
UL1778 APPROVAL: All the UPS systems described in this manual are UL1778 listed. All meet strict requirements for UPS protection. They provide battery backup during blackouts, switching between utility power and battery backup power in 2-4 milliseconds. In addition, all have hospital-grade plugs and receptacles that reduce the possibility of accidental equipment disconnection. All are suitable for use in hospitals, outside patient care areas.
UL2601-1 APPROVAL: Several of the UPS models described in this manual are UL 2601-1 listed (see specifications). They incorporate Faraday-shielded isolation transformers that reduce the cumulative leakage current of all connected equipment to below 300 microamps. These models are suitable for use in hospitals, including inside patient care areas.
FCC CLASS A RADIO/TV INTERFERENCE NOTICE: Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy, and if not installed and used in accordance with the instruction manual, may cause interference to radio communications. Operation of this equipment is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense. The user must use shielded cables and connectors with this product. Any changes or modifications to this product not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
FCC CLASS B RADIO/TV INTERFERENCE NOTICE (U.S. ONLY): Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy, and if not installed and used in accordance with the instruction manual, may cause interference to radio communications. However, there is no guarantee that inter- ference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference using one or more of the following measures: reorient or relocate the receiving antenna; increase the separation between the equipment and the receiver; connect the equipment into an outlet on a circuit different from that which the receiver is connected; consult the dealer or an experienced radio/television technician for help. The user must use shielded cables and connectors with this product. Any changes or modifications to this product not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This device complies with part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.
FCC PART 68 CONSUMER INFORMATION AND FCC REQUIREMENTS (U.S.A. ONLY):
1.This equipment complies with Part 68 of the FCC rules. On the top or bottom of this equipment is a label that contains, among other information, the FCC registration number for this equipment. If requested, provide this information to your telephone company.
2.If your Fax/Modem Protector causes harm to the telephone network, the telephone company may temporarily discontinue your service. If possible, they will notify you in advance. But if advance notice isn't practical, you will be notified as soon as possible. You will be advised of your right to file a complaint with the FCC.
3.Your telephone company may make changes in its facilities, equipment, operations or procedures that could affect the operation of your equipment. If they do, you will be given advance notice so as to give you an opportunity to maintain uninterrupted service.
4.If you experience trouble with this Fax/Modem Protector, please contact Tripp Lite Customer Support, 1111 W. 35th Street, Chicago, Illinois 60609; Phone (773) 869-1234 for repair/warranty
information. The telephone company may ask you to disconnect this equipment from the network until the problem has been corrected or you are sure that the equipment is not malfunctioning.
5.This equipment may not be used on coin service provided by the telephone company. Connection to party lines is subject to state tariffs. (Contact your state public utility commission or corporation for information.)
EQUIPMENT ATTACHMENT LIMITATIONS (MODELS WITH THE INDUSTRY CANADA LABEL IN CANADA ONLY):
Notice: The Industry Canada label identifies certified equipment. This certification means that the equipment meets the telecommunications network protective, operational and safety requirements as prescribed in the appropriate Terminal Equipment Technical Requirements Document(s). The Department does not guarantee the equipment will operate to the user’s satisfaction.
Before installing this equipment, users should ensure that it is permissible to be connected to the facilities of the local telecommunications company. The equipment must also be installed using an acceptable method of connection. The customer should be aware that the compliance with the above conditions might not prevent degradation of service in some situations.
Repairs to certified equipment should be coordinated by a representative designated by the supplier. Any repairs or alterations made by the user to this equipment, or equipment malfunctions, may give the telecommunications company cause to request the user to disconnect the equipment.
Users should ensure for their own protection that the electrical ground connections of the power utility, telephone lines and internal metallic water pipe system, if present, are connected together. This precaution may be particularly important in rural areas. Caution: Users should not attempt to make connections themselves, but should contact the appropriate electric inspection authority, or electrician, as appropriate.
Note on Labeling
Two symbols are used on the label.
V~ : AC Voltage
V: DC Voltage