Anritsu S820A manual Frequency Accuracy, Procedure

Page 2

 

Coupler

Coupler

Test

 

Multiplier

 

 

 

Port

 

 

 

RF

Wave

 

 

Source

 

 

 

Module

 

 

 

Multiplier

 

 

Local

 

 

 

Oscillator

 

 

 

 

Incident

 

Reflected

 

 

 

Local

 

 

 

Oscillator

 

 

 

 

 

Synchronous

 

 

 

Detector

 

Display

A/D Converter

CPU Serial Port

Figure 2. Site Master Block Diagram

4. FREQUENCY ACCURACY

The following test can be used to verify the CW fre- quency accuracy of the Site Master. Measurement calibration of the Site Master is not required for this test.

a.Equipment Required:

￿Spectrum Analyzer Anritsu Model MS2602A

b.Procedure:

1.Press and hold the ESCAPE/CLEAR key, then press the ON/OFF key to turn on the Site Master. (This sets the instrument to the factory preset state.)

NOTE

Before continuing, allow a five minute warm up for the internal circuitry to stabilize.

2.Press the FREQ soft key.

3.Press the F1 soft key, set to 9 GHz, then press the ENTER key.

4.Press the F2 soft key, set to 9 GHz, then press the ENTER key.

5.Connect the RF cable from the Site Mas- ter Test Port to the RF Input on the MS2602A.

6.Set up the Spectrum Analyzer as follows:

(a)Press Preset.

(b)Press Center and enter 1 GHz.

(c)Press the Max Hold button.

7.If the Site Master has gone into the hold mode, press the RUN/HOLD key to make the measurement.

8.Use the Spectrum Analyzer marker to measure the center of the response. The frequency should be 1 GHz ￿75 kHz.

NOTE

Nominal power is approximately 0 - 10 dBm at 1 GHz.

2

Site Master S820A MM

Image 2
Contents Introduction Performance Verification DescriptionProcedure Frequency AccuracyReturn Loss Verification Power Monitor Verification 560-7N50B RF DetectorXXA Precision Return Loss Test Setup Termination VerificationExample of a Good Connection Battery Pack Removal Site Master Battery RemovalBattery Pack Replacement Battery Disposal KEY PAD Membrane Replacement LCD ReplacementMain PCB and RF Module Assembly Replacement Replaceable Parts Replaceable parts for the Site Master are listed in TableRF Measurement Chart Return Loss Verification Worksheet Anritsu Service Centers