Anritsu S820A manual Termination Verification, XXA Precision Return Loss Test Setup

Page 5

 

10. Connect the output of the MA2418A Ref-

b.

Procedure

 

 

erence Source to the two attenuators so

 

 

 

 

 

 

as to add 40 dB of attenuation (Figure 3).

 

 

1. Connect the test equipment as shown in

 

11. Connect the MA2418A Reference Source

 

 

 

Figure 3.

 

 

 

 

 

 

 

and the attenuators to the input of the

 

 

2. Press the Power key on the 541XXA to

 

 

560-7N50B RF detector.

 

 

 

On.

 

12. Verify that the power monitor reading is

 

3.

Press the System Menu key.

 

 

now –40.0 dBm ±2 dB.

 

 

4. Using the Menu up-down keys: Highlight

 

 

 

 

 

 

 

 

 

7.

TERMINATION VERIFICATION

 

 

 

RESET, then press the Select key.

 

 

 

 

This test can be used to verify the accuracy of termi-

 

 

5. At the RESET MENU display, use the

 

 

 

Menu up-down keys to highlight

nations used with the Site Master. The test uses the

 

 

 

RESET TO FACTORY DEFAULTS, then

precision return loss mode of the 541XXA Scalar

 

 

 

press the Select key.

Measurement System. Measurements of termina-

 

6.

Set the signal source for the frequency

tions using this mode provide results that are trace-

 

 

 

 

range as follows:

able to the NIST (National Institute of Standards

 

 

 

 

 

 

 

and Technology) standards for the precision airline.

 

 

 

(a) Press the Frequency key.

a.

Equipment Required:

 

 

 

(b) Using the Data Entry Keypad or

 

 

 

Data Entry Knob, set the START fre-

 

 

 

 

 

 

 

 

 

 

 

￿

Scalar Measurement System,

 

 

 

quency to 0.01 GHz. Press the Enter

 

 

 

 

key.

 

 

Anritsu 54161A/69A

 

 

 

 

 

 

 

 

 

 

￿

Offset SWR Autotester,

 

 

 

(c) Using the Data Entry Keypad or

 

 

 

 

Data Entry Knob, set the STOP fre-

 

 

Anritsu 560-98KF50-15

 

 

 

 

 

 

 

 

quency to 20.0 GHz. Press the Enter

 

 

 

 

 

 

 

 

 

 

 

￿ Precision Airline, Anritsu 19KF50

 

 

 

key.

 

￿

Open/Short, Anritsu 22K50

 

 

7. Press the Channel 2 Display On/Off key

 

￿ 50 Ohm Termination, Anritsu 28K50

 

 

 

to Off.

 

 

 

 

 

 

￿ Source Adapter, Anritsu K220B

 

8.

Press the Channel 1 Menu key.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Open/Short

 

 

 

 

 

 

50 Ohm

 

 

 

 

 

 

Termination

 

 

 

 

 

 

 

 

 

 

 

 

 

Termination

 

 

 

 

 

 

under Test

 

 

 

 

 

 

 

 

 

 

 

 

541XXA Scalar Measurement System

A B

Source

Adapter

Beadless

End

Precision

Air Line

Connect Dashed Line Connections

When Directed By Procedure

Offset SWR

Autotester

Figure 4. 541XXA Precision Return Loss Test Setup

Site Master S820A MM

5

Image 5
Contents Description Introduction Performance VerificationFrequency Accuracy ProcedureReturn Loss Verification 560-7N50B RF Detector Power Monitor VerificationTermination Verification XXA Precision Return Loss Test SetupExample of a Good Connection Site Master Battery Removal Battery Pack RemovalBattery Pack Replacement Battery Disposal LCD Replacement KEY PAD Membrane ReplacementMain PCB and RF Module Assembly Replacement Replaceable parts for the Site Master are listed in Table Replaceable PartsRF Measurement Chart Return Loss Verification Worksheet Anritsu Service Centers