Agilent Technologies E1470A user manual Table B-10. Writing to Relay Control Registers

Page 48

Table B-10. Writing to Relay Control Registers

To connect CH000 to:

COM00

COM01

COM02

COM03

COM04

COM05

COM25

 

 

 

 

 

 

 

 

Write to register 20h

0

132

2244

-29500

-13116

-13116

-13116

 

 

 

 

 

 

 

 

Write to register 28h

x

x

x

x

4

38

-32714

 

 

 

 

 

 

 

 

To connect CH001 to:

COM00

COM01

COM02

COM03

COM04

COM05

COM25

 

 

 

 

 

 

 

 

Write to register 20h

1

133

2245

-29499

-13115

-13115

-13115

 

 

 

 

 

 

 

 

Write to register 28h

x

x

x

x

4

38

-32714

 

 

 

 

 

 

 

 

To connect CH002 to:

COM00

COM01

COM02

COM03

COM04

COM05

COM25

 

 

 

 

 

 

 

 

Write to register 20h

2

134

2246

-29498

-13114

-13114

-13114

 

 

 

 

 

 

 

 

Write to register 28h

x

x

x

x

4

38

-32714

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

To connect CH010 to:

COM01

COM02

COM03

COM04

COM05

COM25

 

 

 

 

 

 

 

 

 

Write to register 20h

0

2112

-29632

-13248

-13248

-13248

 

 

 

 

 

 

 

 

 

Write to register 28h

x

x

x

4

38

-32714

 

 

 

 

 

 

 

 

 

To connect CH011 to:

COM01

COM02

COM03

COM04

COM05

COM25

 

 

 

 

 

 

 

 

 

Write to register 20h

16

2128

-29616

-13232

-13232

-13232

 

 

 

 

 

 

 

 

 

Write to register 28h

x

x

x

4

38

-32714

 

 

 

 

 

 

 

 

 

To connect CH012 to:

COM01

COM02

COM03

COM04

COM05

COM25

 

 

 

 

 

 

 

 

 

Write to register 20h

32

2144

-29600

-13216

-13216

-13216

 

 

 

 

 

 

 

 

 

Write to register 28h

x

x

x

4

38

-32714

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

To connect CH030 to:

COM03

COM03

COM05

COM25

 

 

 

 

 

 

 

 

 

 

 

Write to register 20h

0

16384

16384

16384

 

 

 

 

 

 

 

 

 

 

 

Write to register 28h

x

4

38

-32714

 

 

 

 

 

 

 

 

 

 

 

To connect CH031 to:

COM03

COM03

COM05

COM25

 

 

 

 

 

 

 

 

 

 

 

Write to register 20h

4096

20480

20480

20480

 

 

 

 

 

 

 

 

 

 

 

Write to register 28h

x

4

38

-32714

 

 

 

 

 

 

 

 

 

 

 

To connect CH032 to:

COM03

COM03

COM05

COM25

 

 

 

 

 

 

 

 

 

 

 

Write to register 20h

8192

24576

24576

24576

 

 

 

 

 

 

 

 

 

 

 

Write to register 28h

x

4

38

-32714

 

 

 

 

 

 

 

 

 

 

 

48 Register-Based Programming

Appendix B

Image 48
Contents E1470-90002 Page Contents Appendix C RF Switch Error Messages Index Agilent Technologies Warranty Statement Government Restricted RightsSafety Symbols Documentation HistoryManufacturer’s Name Manufacturer’s AddressDeclares, that the product Product Name Model NumberPage Using This Chapter Switching DiagramChapter Cascade Relays COM 33 332 Configuring the RF Switch Chapter Creating Multiple MultiplexersRF Switch Configuration Maximum POWER. The maximum RF power that can be appliedChapter Configuring the RF Switch On this moduleSetting the Logical Address Setting the Interrupt Request Level Connecting User Wiring Installing User Wiring Configuring the RF Switch ChapterAssembling SMB Connector Jacks SMB Connectors and Connector HousingsUser Connections Wiring Installing Device Drivers RF SwitchAddressing the Switch Programming the RF Switch ChapterChapter Programming the RF Switch Programming ExamplesProgramming the RF Switch Chapter Example Closing a Signal Path Example Opening Closing Signal Paths Example Saving Recalling Module States Programming the RF Switch Chapter Command Types Common Commands FormatScpi Commands Format Chapter RF Switch Command ReferenceScpi Command Reference Linking CommandsRF Switch Command Reference Chapter DIAGnostic DIAGnosticCLOSeDIAGnosticCLOSe? DIAGnosticOPENDIAGnosticOPEN relay,relay... opens individual relays on Reset to its power-on state Common to NCDIAGnosticOPEN? DIAGOPEN333DIAGnosticRELAY? RSTDIAGThis programCLOS042,043,053,054,256REL?returnsROUTe ROUTePATHCOMMonExample Closing Channel Path 10002 SYSTemERRor? SYSTemSYSTemVERSion? Ieee 488.2 Common Commands Quick Reference CommandCommands Quick Reference DescriptionAppendix a RF Switch Specifications Appendix aRF Switch Specifications Appendix a Appendix B About This AppendixRegister Addressing Addressing OverviewBase Address Figure B-1. Register Address Locations Within VXI A16A16 Address Space Outside the Command Module Register Definitions Table B-1. General Register-based Programming MethodTable B-2. Register Map Register-Based Programming Appendix BTable B-3. Status Register Bit Patterns read Table B-4. Control Register Bit Pattern writeRelay Control Registers Vxie Table B-10. Writing to Relay Control Registers Register Programming Example Create and open a device sessionWait 5 seconds before resetting module Void errhandler /* Error Handling Routine Register-Based Programming Appendix B Appendix C RF Switch Error Messages Appendix CRF Switch Error Messages Appendix C Index IndexIndex E1470-90002