Agilent Technologies E1470A user manual RF Switch Configuration, On this module

Page 13

COM 04 can be used for a 15-to-1 multiplexer for all channels between 000 and 042. COM 05 can be the common for all channels from 000 through 052 creating an 18-to-1 multiplexer. Multiplexers of 21-to-1, 24-to-1, and 27-to-1 can also be configured. Two 30-to-1 multiplexers can be created using channels 00 through 132 to COM 05 and channels 200 through 332 to COM 25. One 60-to-1 multiplexer can be created using all the channels to COM 25.

RF Switch Configuration

This section gives guidelines to configure the RF Switch module, including:

Warnings and Cautions

Selecting the Logical Address

Setting the Interrupt Request Level

Connecting User Wiring

Warnings and

Cautions

WARNING

SHOCK HAZARD. Only service-trained personnel who are

 

aware of the hazards involved should install, remove, or

 

configure the module. Before you remove any installed

 

module, disconnect AC power from the mainframe and from

 

other modules that may be connected to the module.

 

 

 

 

WARNING

CHANNEL WIRING INSULATION. All channels that have a

 

common connection must be insulated so that the user is

 

protected from electrical shock in the event that two or more

 

channels are connected together. This means wiring for all

 

channels must be insulated as though each channel carries

 

the voltage of the highest voltage channel.

 

 

 

 

CAUTION

MAXIMUM POWER. The maximum RF power that can be applied

 

to the module is 10 Watts RF. Do not apply line AC power to any terminal

 

on this module.

 

 

 

 

CAUTION

STATIC ELECTRICITY. Static electricity is a major cause of component

 

failure. To prevent damage to the electrical components in the module,

 

observe anti-static techniques whenever removing a module from the

 

mainframe or working on a module.

 

 

Chapter 1

Configuring the RF Switch 13

Image 13
Contents E1470-90002 Page Contents Appendix C RF Switch Error Messages Index Government Restricted Rights Agilent Technologies Warranty StatementDocumentation History Safety SymbolsManufacturer’s Address Manufacturer’s NameDeclares, that the product Product Name Model NumberPage Switching Diagram Using This ChapterChapter Cascade Relays COM 33 332 Creating Multiple Multiplexers Configuring the RF Switch ChapterMaximum POWER. The maximum RF power that can be applied RF Switch ConfigurationChapter Configuring the RF Switch On this moduleSetting the Logical Address Setting the Interrupt Request Level Installing User Wiring Configuring the RF Switch Chapter Connecting User WiringSMB Connectors and Connector Housings Assembling SMB Connector JacksUser Connections Wiring RF Switch Installing Device DriversProgramming the RF Switch Chapter Addressing the SwitchProgramming Examples Chapter Programming the RF SwitchProgramming the RF Switch Chapter Example Closing a Signal Path Example Opening Closing Signal Paths Example Saving Recalling Module States Programming the RF Switch Chapter Common Commands Format Command TypesScpi Commands Format Chapter RF Switch Command ReferenceLinking Commands Scpi Command ReferenceRF Switch Command Reference Chapter DIAGnosticCLOSe DIAGnosticDIAGnosticOPEN DIAGnosticCLOSe?DIAGnosticOPEN relay,relay... opens individual relays on Reset to its power-on state Common to NCDIAGOPEN333 DIAGnosticOPEN?RSTDIAGThis programCLOS042,043,053,054,256REL?returns DIAGnosticRELAY?ROUTePATHCOMMon ROUTeExample Closing Channel Path 10002 SYSTem SYSTemERRor?SYSTemVERSion? Command Ieee 488.2 Common Commands Quick ReferenceDescription Commands Quick ReferenceAppendix a Appendix a RF Switch SpecificationsRF Switch Specifications Appendix a About This Appendix Appendix BRegister Addressing Addressing OverviewFigure B-1. Register Address Locations Within VXI A16 Base AddressA16 Address Space Outside the Command Module Table B-1. General Register-based Programming Method Register DefinitionsTable B-2. Register Map Register-Based Programming Appendix BTable B-4. Control Register Bit Pattern write Table B-3. Status Register Bit Patterns readRelay Control Registers Vxie Table B-10. Writing to Relay Control Registers Create and open a device session Register Programming ExampleWait 5 seconds before resetting module Void errhandler /* Error Handling Routine Register-Based Programming Appendix B Appendix C Appendix C RF Switch Error MessagesRF Switch Error Messages Appendix C Index IndexIndex E1470-90002