VXI SM8000 Interrupt Status Register Read Only, Interrupt Control Register Read and Write

Page 41

VXI Technology, Inc.

Interrupt Status Register – Read Only

ADDR

Plug-In LA+0x1A

 

D15

Scan Function

The latest scan list update is complete.

done

 

 

D14

Openbus Active

The Openbus was activated by one or more programmed inputs.

Event true

See description of the Openbus in the module register section.

 

 

 

D13 = Module 5, … and D8 = Module 0.

D13-D8

Modules 0-5 Busy

The programmed Busy signal from one of the modules has timed

complete

 

out. This indicates that the relays actuated for that BUSY cycle

 

 

 

 

have settled and a measurement may take place.

D7-D0

Reserved

Always reads back as FFFF16

Note: This status register may be used in a polled fashion rather than allowing the events above to generate an Interrupt. A read of this register will clear any active bits. Bits that are not set, or are about to be set, are not affected by a read of this register.

Interrupt Control Register – Read and Write

ADDR

Plug-In LA+0x1C

 

 

D15

Scan Function

0

= enabled

done mask bit

1

= disabled

 

D14

Openbus Active

0

= enabled

Event true mask bit

1

= disabled

 

 

Module 0-5 Busy

0

= enabled

D13-D8

1

= disabled

complete

 

 

 

D13 = Module 5 … and D8 = Module 0.

 

 

D7

IR ENA*

0

= writing a 0 to this bit enables interrupter capabilities

1

= writing a 1 to this bit disables interrupter capabilities

 

 

D6

IH ENA*

The module has no interrupt handler capability, therefore writing a

1 or 0 has no effect. A 1 is always read back for this bit.

 

 

 

Interrupter IRQ

The complement of the value programmed into these three bits

D5-D3

reflects the selected IRQ line used by the module. A value of 0112

Line

would select IRQ4, a value of 0002 would select IRQ7, and a

 

 

 

value of 1112 would disconnect the IRQ lines.

 

 

The module has no interrupt handler capability; therefore writing

D2-D0

Handler IRQ Line

to these bits has no effect. A 1112 is always read back for these

 

 

bits.

Note: That all bits in this register are set to 1s upon receipt of a hard or soft reset.

 

 

Subclass Register – Read Only

 

ADDR

Plug-In LA+0x1E

 

 

D15

VXIbus Extended

Always reads as 1.

 

Device

 

 

 

 

D14-D0

Extended Memory

Always reads as 7FFD16

 

Device

 

 

 

SM8000 Series Programming

41

Image 41
Contents SM8000 Series VXI Technology, Inc Table of Contents Control Modes VXI Technology, Inc SM8000 Series Preface Certification Steve Mauga, QA Manager EMCVXI Technology, Inc SM8000 Series Preface Terms and Symbols Service should only be performed by qualified personnelAvoid Electric Shock Support Resources VXI Technology, Inc SM8000 Series Preface Overview SectionSM8000 Series Optical Switch Controller Configurations SM8001 / SM8002 MULTI-CHANNELSWITCHESDuplex 1 x N PDL6 SM8001 / SM8002 Multi Switch SpecificationsSpdt SM8003 Prism SwitchesPDL3 SM8003 Prism Switch SpecificationsPDL SM8101 / SM8102 Optical AttenuatorsIntroduction Calculating System Power and Cooling RequirementsSetting the Chassis Backplane Jumpers MSB LSB Setting the Logical AddressExample Selecting the Extended Memory Space Optical Connections Cleaning Optical ConnectorsMating Optical Connectors General Description SM8001 / SM8002 Multi-Channel SwitchesSM8003 Prism Switches SM8003 Prism SwitchesSM8101 / SM8102 Optical Attenuators Attenuator DiagramRelay Registers Output Channel Selection Resetting the SwitchOperation N Switch Configuration Common Active Channel Duplex 1 x N Switch ConfigurationN Blocking Switch Configuration N Non-Blocking Switch Configuration 8MULTI-SWITCH Timing Calculating Switching TimeSM8003 Prism Switches Control Modes Starting the DeviceUncalibrated Operation Move-To-Absolute-Step Calibrated OperationError Status Busy SignalResetting the Device Commanding the DevicesAddressing Register AccessOffset Write Function Smip II Register MAP A16Logical Address Register Write Only ID Register Read OnlyDevice Type Register Read Only Status Register Read OnlyOffset Register Read and Write Control Register Write OnlySerial Number High Register Read Only Serial Number Low Register Read OnlySubclass Register Read Only Interrupt Status Register Read OnlyInterrupt Control Register Read and Write NVM Access Register Write NVM Access Register ReadBoard X, Y Used Address Register Read and Write Trace RAM Start High Register Read and WriteTrace RAM End Low Register Read and Write Trace RAM End High Register Read and WriteTrace RAM Address High Register Read and Write Trace RAM Address LOW Register Read and WriteTTL Trigger Polarity Register Write Only Open Trigger Select Register Write OnlyBusy Trigger Control Register Read and Write Trace RAM Control Register Read and WriteReserved Registers Read and Write Trigger Advance Register Write OnlyBoard Busy Register Read Only 1MB RAM Control Register Read and Write See Typical Optical Multi Switch OperationControl Register Delay Register Read and Write Plug-In LA+0x104 Command Register Write Only Address Register Write Only Relay Register Offset Device MemoryWriting to the Relay Registers Relay Register 00 Read and WriteRegister 0A thru 0C Read Relay Optical Module’s Data Attenuation LevelRegister 02 thru 08 Read and Write Programming Examples Typical Optical MULTI-SWITCH Control ExampleWrite Typical Optical Attenuator Control ExampleVXI Technology, Inc Read Example Write ExampleCommand Transmit Data Receive Data Byte Byte Count Command SET31h 30h35h 32h80h 6Ch82h 81h89h 83h8Bh 8Ah8Dh 8Ch8Eh Convert to hexadecimal 90hRecommends That the Optical MODULE’S A2h 96hVXI Technology, Inc SM8000 Series Programming Index VXIbus

SM8000 specifications

The VXI SM8000 is a versatile and robust modular instrumentation platform that is designed for various testing applications in industries such as aerospace, telecommunications, and military. This advanced test equipment is part of the VXI (Versatile Instrumentation System) standards, which allow for high-speed data acquisition, processing, and analysis.

One of the main features of the VXI SM8000 is its modular architecture, which enables users to configure the system according to their specific testing needs. This flexibility means that users can integrate various modules, including signal generators, oscilloscopes, and digital multimeters, making the system adaptable to a wide range of applications. The modular design also ensures scalability, allowing for easy expansion as test requirements grow.

The VXI SM8000 supports high-speed communication via its advanced backplane design. This design features multiple data buses that facilitate fast communication between modules, ensuring minimal latency and enhanced performance during testing. Moreover, the system is built to accommodate high-density modules, maximizing the amount of test parameters that can be handled simultaneously.

The SM8000 boasts enhanced signal processing capabilities thanks to its digital signal processing (DSP) technology. This feature enables real-time analysis and processing of signals, which is crucial for applications that require rapid feedback and adjustments. Additionally, the system is equipped with high-resolution analog-to-digital converters, ensuring precise measurement and analysis.

One of the key characteristics of the VXI SM8000 is its reliability. Built with robust materials and advanced engineering, the system is designed to endure rigorous testing environments and extreme operating conditions. This durability ensures consistent performance over time, making it a reliable choice for critical applications.

Furthermore, the VXI SM8000 is compatible with various software platforms, enabling seamless integration into existing testing environments. It supports programming languages like LabVIEW and MATLAB, allowing engineers to develop customized test scripts and configurations easily.

In summary, the VXI SM8000 is a powerful and flexible test system that combines modularity, high-speed communication, advanced signal processing, and compatibility with industry-standard software. Its reliability and scalability make it an ideal choice for professionals seeking a comprehensive testing solution across various sectors.