Perle Systems TS2, DS1 Calibrating Temperature Input, Calibrating Thermocouple, Calibrating RTD

Page 194

Calibrating I/O

Calibrating Temperature Input

To calibrate an Analog (Temperature) input channel, read the section that applies to the type of input you are calibrating. Note that calibration will be done for the active channel configuration; for example, if Channel A1 is set to thermocouple, you cannot calibrate it for RTD. During the calibration process, you will be asked to apply the minimum and maximum range value to the channel in either mV or Ohms; for example, to calibrate for thermocouple J 0 to 760C, you will be prompted to first apply -80mV and then +80mV to the channel.

Also, you cannot actively calibrate disabled channels (although if you enable the channel and then set it for the type of thermocouple or RTD that has already been calibrated on another channel, it will also be calibrated).

Calibrating Thermocouple

When calibrating the IOLAN Analog input for thermocouple, you will need a calibration meter that is better than .15% accuracy. When you calibrate one channel, all thermocouple channels are automatically calibrated for that range; if another channel is set for a different range, you will need to calibrate that channel separately, but all channels that use that range are automatically calibrated.

Calibrating RTD

When calibrating the IOLAN Analog input for RTD, you will need a resistor that is better than .05% Ohms accuracy. When you calibrate one channel, all RTD channels are automatically calibrated for that range; if another channel is set for a different range, you will need to calibrate that channel separately, but all channels that use that range are automatically calibrated.

194

Image 194
Contents Iolan DS1/TS2 EN 55022 1998, Class A, Note Intended Audience PrefaceAbout This Book DocumentationTypeface Conventions Online HelpTable of Contents Configuration Methods Getting Started Using DeviceManager and WebManager Serial Port Profiles Network Settings Configuring Users Configuring Security UDP Configuring the System Controlling the I/O Channels Appendix a Virtual Modem AT Commands Appendix D I/O Wiring Diagrams Appendix H Data Logging Appendix I Modbus Remapping 241 Iolan Family Models IntroductionAbout the Iolan Iolan Features HardwareSoftware Security Iolan Components ConnectivityIntroduction Desktop Models Power Supply SpecificationsGetting to Know Your Iolan OverviewDS1 Serial ActivityTop View End ViewSerial Mode Console/Serial SwitchConsole Mode Models with Terminal Block Connector Powering Up the IolanModels with Barrel Connector Configuration Methods Chapter Requires a Configured IP Address Configuration Methods OverviewConfigures an IP Address Easy Config Wizard Connecting to the Iolan Using DeviceManager Access PlatformsFeatures DeviceManagerDeviceManager Using DeviceManager Navigating the OptionsWebManager Connecting to the Iolan Using WebManagerDownloading the Configuration WebManager Using WebManager Through the Network Command Line InterfaceConnecting to the Iolan Using the CLI Menu Using the CLIConnecting to the Iolan Using the Menu Through the Serial PortUsing the Menu Connecting to the Iolan Using DHCP/BOOTP Using DHCP/BOOTPDHCP/BOOTP Parameters Connecting to the Iolan Using Snmp SnmpUsing the Snmp MIB Easy Configuration Wizard Getting StartedSetting Up the Network Using a Direct Serial Connection to Specify an IP Address Using a Direct Serial Connection to Enable BOOTP/DHCP Using ARP-Ping For an IPv6 NetworkSetting Up the Serial Ports Setting Up the Serial Ports Setting Up Users Using DeviceManager WebManager Navigating DeviceManager/WebManager DeviceManagerWebManager Using DeviceManager to Connect to the Iolan Starting a New SessionAssigning a Temporary IP Address to a New Iolan Adding/Deleting IOLANs Manually Logging in to the IolanCreating a New Iolan Configuration in DeviceManager Using WebManager to Connect to the IolanConfiguration Files Logging into the IolanManaging the Iolan Opening an Existing Configuration FileImporting an Existing Configuration File Serial Ports FunctionalityEditing a Serial Port Resetting a Serial Port Overview Serial Port ProfilesCommon Tabs Hardware Tab Field DescriptionsDefault Auto Default NoneDefault Full Enable InboundPacket Forwarding Tab Field Descriptions Packet Definition Enable Trigger1Enable Trigger2 Packet SizeConsole Management Profile FunctionalityProtocol Listen for Connections on TCP Port Default TelnetGeneral Tab Field Descriptions IP AddressAdvanced Tab Field Descriptions Default 10 ms Dial Dial OutDial Retry Dial TimeoutTruePort Profile ModemConnect to remote System Default Enabled Host NameConnect Default Disabled Define Additional TCP PortAdding/Editing Additional TruePort Hosts Primary Host Backup HostAdding/Editing a Multihost Entry When Signals highDay Motd Default Disabled Enable TCP Enable Data Logging Trueport Lite Mode Dial Retry Connection Workstation/Server on the network Default Enabled TCP Sockets ProfileDefault Disabled Adding/Editing Additional Hosts Adding/Editing a Multihost Entry Day Motd Default Disabled Enable Data LostDefault 10 ms Dial UDP Sockets Profile Serial Port Profiles UDP Entry Start IP Address Connections onUDP port Direction End IP AddressTerminal Profile Port UDP port to use Default 0 zeroDefault 10 ms Session StringsTVI925 Default DumbVT100 VT320 specifically supporting VT320-7Automatically Default Disabled ProtocolWhen any data is ReceivedData Range Default Disabled Hotkey PrefixSession Timeout User Service Settings Login SettingsQuit Telnet SettingsInterrupt EraseSerial Tunneling Profile Server Client Default Disabled Host NameAct As Tunnel Delay After Break Break LengthVirtual Modem Profile 100 DTR Signal Acts as RTS Signal Always On DCDResponse Delay RTS Signal Acts asAT Command Phone Number to Host Mapping VModem Phone Number EntryHost IP Address Control Signal I/O ProfilePhone Number Auto Clear Mode Input Signal Field DescriptionsInvert Signal DescriptionManual Clear Output Signal Field DescriptionsMode Default Disabled Syslog Failsafe ActionModbus Gateway Profile IP Mappings IP address Advanced Slave Settings ButtonDestination Slave ButtonDefault 1000 ms Default 30 msMessage Timeout Advanced Field DescriptionsUID Start Modbus Slave IP Settings Field DescriptionsAdding/Editing Modbus Slave IP Settings UID End Default HostDefault TCP TypeModbus Slave Advanced Settings Field Descriptions General Tab Field Description Default OffCustom Application Profile Advanced Advanced Serial Settings TabClosing Serial Port Default Disabled Deny Multiple Field DescriptionsModems Tab Adding/Editing a ModemName Name of the modem TruePort Baud Rate Tab Field DefinitionsNetwork Settings System Name IP SettingsIPv4 Settings DomainIPv6 Settings Adding/Editing a Custom IPv6 Address Default IPV6 68-1500 bytes AdvancedAdd Button Adds a host to the host table Edit Button Host TableAdding/Editing a Host Delete Button Deletes a host from the host tableRoute List Adding/Editing RoutesIPv6 Prefix Bits Default DefaultIPv4 Subnet Mask HostConfiguring Users User Settings Adding/Editing Users General TabDefault Normal LevelDefault DSPrompt Services TabService Host IPLanguage Default EnglishAdvanced Tab Sessions Tab Session 1, 2, 3 ConnectSerial Port Access Tab Configuring Security ServicesField Descriptions Configuring I/O Interfaces Technical SpecificationsSettings Access FunctionalityAdvanced Slave Modbus Settings Failsafe Timer Functionality UDP Functionality Default Celsius Temperature FunctionalityUDP Settings UDP EntryMonitoring Application A4D2 ChannelsAnalog Analog I/O AnalogDefault Current Alarm SettingsMonitoring Application Digital I/OInput Mode 147 Output Mode Default SinkInactive Signal Default ManualPulse Mode Active SignalMonitoring Application A4D2Relay Relay I/O151 Warehouse Perle Iolan Reception Digital I/O ExtensionFront Door Remote HostInput number 1 Byte Current Alarm State 1 Byte z 0 = Not in alarmMessage type 1 Byte Input TCP Port Enable I/OExtension Hosts Local connectionSimultaneously Communicate to the Primary Host Temperature Temperature I/O Industrial FreezersDefault RTD Alarm Settings Basic Analog Alarm SettingsTrigger Type Advanced Analog Alarm SettingsClear Mode ClearUDP Unicast Format UDP Broadcast PacketSection Channel Length EnabledAnalog Section Length Channel Serial Pin Signal SectionDigital/Relay Section Length PinModbus Serial Application Connected to the Serial Port Modbus SlaveUDP Unicast Example Modbus Serial Application Connected to the NetworkFunction Codes Modbus I/O AccessModbus TCP Application Coil/Register Descriptions Data Model A1/T1 A2/T2 A3/T3 A4/T4 Holding Registers Serial Port Coil/Register DescriptionsA4/T4 Registers Input RegistersA4D2/A4R2 Registers Data Model D1/R1 D2/R2 CoilsData Model D3/R1 D4/R2 Coils Serial Pin SignalsD4/D2R2 Registers Data Model Pin CoilsTruePort/Modbus Combination Modbus Serial Application TruePort Power Digital OutputTruePort I/O PC running aAPI Over TruePort Only PC running Custom Application API TruePortIntroduction Accessing I/O Data Via TruePortSetup Command Format Format of API CommandsGet Commands Response FormatSet Commands Example 1 Turn on the first relay on a D2R2 unit Successful Response FormatUnsuccessful Response Format Example 2 Turn on the first and second relay on a D2R2 unitError Codes Snmp TrapsSyslog Configuring the System ChapterAlerts Management Snmp Traps Tab Field Descriptions Field Description Default 1 secondCustom App/Plugin Login Tab Field Descriptions Bootup Files Tab Field Descriptions Tftp Host WebManager WebManager. Default DisabledMessage of the Day Motd Tab Field Descriptions FilenameRetry Default 3 secondsTftp Tab Field Descriptions Controlling the I/O Channels Activate Output Manually activates the channel output Saving Configuration Files System Administration ChapterManaging Configuration Files Downloading Configuration Files Server Name Downloading Configuration Files to Multiple IOLANsReboot Server Uploading Configuration Files Specifying a Custom Factory Default ConfigurationCalibrating I/O Resetting the Iolan to the Default ConfigurationDownloading Iolan Firmware Calibrating Analog InputCalibrating Voltage Calibrating CurrentCalibrating RTD Calibrating Temperature InputCalibrating Thermocouple Resetting Calibration Data Calibrating Analog ChannelsResetting the Iolan to Factory Defaults Setting the IOLAN’s Date and TimeRebooting the Iolan Language Support Loading a Supplied LanguageSoftware Upgrades and Language Files Translation GuidanceDownloading Terminal Definitions Creating Terminal Definition FilesResetting Configuration Parameters For exampleLost admin Password Commands Virtual Modem Initialization CommandsAT&Cn AT&SnAT&Rn DB25 Male Pinouts and Cabling DiagramsSerial Pinouts DB25 Female RJ45 DB9 Male Serial Only DB9 Male I/ODB25 Male EIA-232 Cabling DiagramsTerminal DB25 Connector DB25 Female10-pin Pin RJ45DB9 Male Modem DB25 Connector Iolan RJ45 DCERxD TxD DTR 20 DTR GND Iolan DB25 Male/Female Setting JumpersPort Iolan Iolan RJ45 DIP Digital I/O Module Analog Input Module Digital Input Dry Contact Wiring I/O DiagramsDigital Input Wet Contact Digital Output Sink Digital Output SourceCurrent Temperature InputAnalog Input VoltageRTD 4-Wire RTD 2-WireRTD 3-Wire Relay OutputNormally Closed Contact TruePort Utility TruePortAPI Response Format API I/O Access Over TruePortAPI Request Format Error Codes Accessories Starter Kit Adapters/CableRJ45F to DB25M DTE Crossover Adapter RJ45F to DB25M DCE Modem Adapter RJ45F to DB25F DTE Crossover Adapter RJ45F to DB9M DTE Crossover Adapter Sun/Cisco RJ45M Connector Cable for Rack Mount Models RJ45F to DB9F DTE Crossover AdapterStarter Kit Adapters/Cable Accessories Communication Issues TroubleshootingHardware Problems Power/Ready LED continues to blink green in desktop modelsLogin Problems DeviceManager ProblemsHost Problems Login Problems Problems with Terminals Unknown IP AddressLong Reboot Cycle DHCP/BOOTP ProblemsLanguage Problems ModelsIPv6 Issues Contacting Technical Support Making a Technical Support QueryRepair Procedure Feedback on this ManualTCP Socket Profile Data LoggingTrueport Profile Create a configuration file Modbus Remapping FeatureConfiguring the Modbus UID Translation Feature Line format for one UID isConfiguring the Modbus UID Translation Feature Glossary Roaming Callback RIP RoutingInformation Protocol Silent ConnectionBootp IndexSnmp
Related manuals
Manual 224 pages 30.96 Kb Manual 88 pages 61.31 Kb Manual 122 pages 55.83 Kb

DS1, TS2 specifications

Perle Systems TS2 and DS1 are advanced serial-to-Ethernet devices designed to facilitate seamless integration of serial devices with modern IP networks. These devices serve as crucial tools for businesses looking to enhance their communication capabilities while leveraging existing infrastructure.

One of the standout features of the Perle TS2 and DS1 is their ability to support multiple serial protocols including RS232, RS422, and RS485. This versatile support allows businesses to connect various legacy devices to their networks, ensuring that critical equipment can still communicate effectively in an increasingly digital landscape.

The TS2 model is particularly noted for its dual serial ports, enabling simultaneous communication with multiple devices. This capability is invaluable for applications requiring real-time data transmission, such as remote monitoring and automated control systems. The DS1 model, on the other hand, is designed for single-port use, making it an ideal solution for smaller-scale operations or specific single-device connections.

Both the TS2 and DS1 incorporate advanced features such as configuration through a web interface, which simplifies setup and management. Users can easily access device settings and monitor performance metrics remotely, enhancing operational efficiency. These devices also support various network protocols such as TCP/IP, UDP, and ICMP, ensuring compatibility with a wide range of network architectures.

In terms of security, Perle Systems prioritize user data protection. The TS2 and DS1 offer features including SSL encryption and SSH protection, providing an added layer of security for sensitive information transmitted over the network. This is particularly important for industries where data integrity and security are paramount, such as healthcare and finance.

Additionally, the rugged design of both devices ensures reliable performance even in harsh environments. With extended temperature ratings and robust housing, the TS2 and DS1 can operate in diverse conditions without compromising functionality.

To summarize, the Perle Systems TS2 and DS1 provide a reliable and flexible solution for connecting legacy serial devices to modern networks. With their support for multiple serial protocols, web-based management, robust security features, and rugged design, these devices cater to a variety of industries looking to modernize their communication infrastructure while maintaining existing equipment. Whether for large-scale operations or smaller implementations, the TS2 and DS1 equip businesses with the tools needed for effective serial communication in an IP-driven world.