Omega Engineering OM-DAQ-USB-2401 manual Measurement Duration, Scan Time, and Resolution

Page 34

4Software

NOTE:

In DAQ applications, thermocouples should not be connected single-ended. Doing so can result in noise and false readings. This is especially true when acquiring other high-amplitude signals in conjunction with thermocouple signals that are connected single-ended.

The DAQ units include built-in cold-junction compensation (CJC), which is automatically invoked when you select TC measurements. The OM-DAQ-USB- 2401 automatically converts acquired voltage readings into compensated- linearized temperature readings. This DAQ system can make thermocouple and volts measurements concurrently.

OM-DAQ-USB-2401

TERMINAL BLOCK

(PARTIAL)

A

A

C

A

A

1

1

O

2

2

H

L

M

H

L

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

V1

V2

COM

V3+

V3

V3-

Figure 4-26. OM-DAQ-USB-2401 Single-Ended and Differential Connections to

Analog Input Channels Diagram

Figure Notes: *V1 and V2 are single-ended inputs.

*V3 is a thermocouple and is a differential input.

4.9 Measurement Duration, Scan Time, and Resolution

In relation to sampling analog input, the terms measurement duration, sample rate, and resolution have the following meanings:

Measurement duration (per channel) – the amount of time used for sampling a channel’s input signal. The measurement durations for the OM-DAQ-USB-2401's analog channels range from very slow (400 milliseconds for one measurement duration) to very fast (1 millisecond for one measurement duration).

Scan Time (per scan) - The amount of time used for scanning (sampling) all selected channels’ input signal. For the OM-DAQ-USB-2401, the scan rates range from very slow (1 hour per scan) to very fast (1 milliseconds per scan).

4-16

Image 34
Contents OM-DAQ-USB-2401 Servicing North America OM-DAQ-USB-2400 Multiple Channel USB Data Acquisition Module OM-DAQ-USB-2400 International Usage and CE Marking 10-1List of Figures SectionExamples of Under Sampling Precautions IntroductionStatement on CE Marking Safety Warnings and IEC SymbolsGeneral Description & System Components System Components Package Inspection HardwareIncluded Items USB Hubs and Power Adaptors Hardware SetupOM-DAQ-USB-2401 System Components Self-powered Hubs USB-powered HubsPower Adaptors Connecting Your OM-DAQ-USB-2401 Acquisition System to PC Connecting Various Hardware Setups Direct Connection to Computer USB PortsConnection to USB-powered Hub Connections to Self-Powered and USB-powered HubsConnections to Self-Powered Hub and to USB-Powered Hub OM-DAQ-USB-2401 MountingOM-DAQ-USB-2401 Wall Mounting OM-DAQ-USB-2400 SeriesOM-DAQ-USB-2401 DIN Rail Mounting OM-DAQ-USB-2401 Rail MountingGetting Started Software InstallationSoftware System RequirementsSoftware Setup Welcome ScreenConfirm Installation Screen DAQ Central Software Operation Main Control WindowToolbar Buttons Toolbar ButtonsMain Control Window Pull-Down Operations File MenuTools Menu 2 Device Configuration Channels TabConfiguration Parameter Settings Manual Acquisition is manually started/stopped Level Data Tab Data Displays18. Main Window Screen 23. Waveform Screen Optical Isolation Power Line RejectionA/D Conversion Analog Input Configuration Input RangesMeasurement Duration, Scan Time, and Resolution Automatic Calibration 27. Calibration/Scan Arrangement28. Thermocouple Wiring Thermocouple MeasurementsResolution Effective Number of Bits ENOB, RMS Signal ManagementScan Time and Resolution Channel Control and ExpansionUnder Sampling and Aliasing RMSSignal Examples of Under SamplingSignal Modes TriggeringNoise Reduction in Differential Mode Averaging System NoiseAnalog Filtering Vadc = VT x Ri / Rs + Ri Input and Source ImpedanceCrosstalk = RCCE Conformity OM-DAQ-USB-2401 Design for CE ConformityFerrite Core Basic Checklist TroubleshootingSymptoms and Solutions Service & Calibration Service & CalibrationSpecifications Thermocouple Input Range Power SupplyExternal Power Source Cold JunctionThermocouple Accuracy ºC for OM-DAQ-USB-2401 TC Type Temp Typical Fast SlowApprovals and Regulatory Compliance International Usage & CE Marking10-2 10-3 10-4 WARRANTY/DISCLAIMER Temperature