Omega Engineering OM-DAQ-USB-2401 manual Signal, Examples of Under Sampling

Page 39

Signal Management 5

The following table provides general advice regarding the selection of Scan Time. The concepts are further illustrated by the figure, Examples of Under Sampling.

 

Scan Time

 

 

 

(Measurement

Scan Rate

 

Analog Input Signal

Duration)

(Sample Rate)

Resolution

 

 

 

 

Steady, or gradual

 

 

 

change

Long

Low

High

 

 

 

 

Highly variable

 

 

 

(unsteady)

Short

High

Low

 

 

 

 

Figure 5-1. Examples of Under Sampling

The above figure depicts three signals from the same temperature fluctuations. Under Sampling (aliasing) is evident in two of the signals. Elaboration follows:

Signal 1:

This signal, based on 1 sample(scan) per division, is represented by a heavy solid line and sample-points designated by polygon symbols. The signal represented is a fairly accurate portrayal of the actual temperature fluctuations.

Signal 2:

This signal, based on 1 sample(scan) every 3 divisions, is represented by a heavy dotted line and sample-points designated by squares. Under sampling has resulted in a distortion, in effect, a lower amplitude than that exhibited by the first signal, even though each measured point is accurate.

Signal 3:

This signal, based on 1 sample(scan) every 4 divisions, is represented by a dotted/dashed line and sample-points designated by plus signs (+). Fewer samples (a greater degree of under sampling) has resulted in a further distortion and lowering of signal amplitude.

From these examples it should be realized that more samples(scans) will result in a more accurate representation of the actual signal, and that under sampling will tend to lower the amplitude of the signal, exhibiting a trend toward a “flat line” state.

As stated earlier, the more variability a signal has, the more samples that are needed to accurately portray it.

5-3

Image 39
Contents OM-DAQ-USB-2401 Servicing North America OM-DAQ-USB-2400 Multiple Channel USB Data Acquisition Module International Usage and CE Marking 10-1 OM-DAQ-USB-2400Section List of FiguresExamples of Under Sampling Introduction PrecautionsSafety Warnings and IEC Symbols Statement on CE MarkingGeneral Description & System Components System Components Hardware Package InspectionIncluded Items Hardware Setup USB Hubs and Power AdaptorsOM-DAQ-USB-2401 System Components USB-powered Hubs Self-powered HubsPower Adaptors Connecting Your OM-DAQ-USB-2401 Acquisition System to PC Direct Connection to Computer USB Ports Connecting Various Hardware SetupsConnections to Self-Powered and USB-powered Hubs Connection to USB-powered HubOM-DAQ-USB-2401 Mounting Connections to Self-Powered Hub and to USB-Powered HubOM-DAQ-USB-2400 Series OM-DAQ-USB-2401 Wall MountingOM-DAQ-USB-2401 Rail Mounting OM-DAQ-USB-2401 DIN Rail MountingSystem Requirements Software InstallationSoftware Getting StartedWelcome Screen Software SetupConfirm Installation Screen Main Control Window DAQ Central Software OperationToolbar Buttons Toolbar ButtonsFile Menu Main Control Window Pull-Down OperationsTools Menu Channels Tab 2 Device ConfigurationConfiguration Parameter Settings Manual Acquisition is manually started/stopped Level Data Displays Data Tab18. Main Window Screen 23. Waveform Screen Power Line Rejection Optical IsolationA/D Conversion Input Ranges Analog Input ConfigurationMeasurement Duration, Scan Time, and Resolution 27. Calibration/Scan Arrangement Automatic Calibration Thermocouple Measurements 28. Thermocouple WiringChannel Control and Expansion Signal ManagementScan Time and Resolution Resolution Effective Number of Bits ENOB, RMSRMS Under Sampling and AliasingExamples of Under Sampling SignalTriggering Signal ModesNoise Reduction in Differential Mode System Noise AveragingAnalog Filtering = RC Input and Source ImpedanceCrosstalk Vadc = VT x Ri / Rs + RiOM-DAQ-USB-2401 Design for CE Conformity CE ConformityFerrite Core Troubleshooting Basic ChecklistSymptoms and Solutions Service & Calibration Service & CalibrationSpecifications Cold Junction Power SupplyExternal Power Source Thermocouple Input RangeTC Type Temp Typical Fast Slow Thermocouple Accuracy ºC for OM-DAQ-USB-2401International Usage & CE Marking Approvals and Regulatory Compliance10-2 10-3 10-4 WARRANTY/DISCLAIMER Temperature