Digi HygroClip DI digital interface instruction manual Overview, Models, HygroClip DI-3

Page 3

IN-E-HyClip DI-V1_10

Rotronic AG

Bassersdorf, Switzerland

Document code

Unit

 

 

 

HygroClip DI digital interface: instruction

Instruction Manual

manual

 

Document Type

Document title

Page 3 of 24

 

 

 

 

 

1.Overview

The HygroClip DI (Digital Interface) allows connecting up to two HygroClip digital humidity- temperature probes to a PC, either by means of a USB port or by means of a wired / wireless TCP/IP network (LAN / Internet). The HygroClip DI can also be used to connect to a PC other instruments from ROTRONIC that have a RS-485 port such as the HygroFlex and M33 transmitters, HygroLab 2 indicator, etc.

The HygroClip DI is designed to operate with the ROTRONIC HW4 networking software (v 1.2.2 or higher). Other than the capacity of the PC, there is no limit to the number of HygroClip DI that can be networked with the PC.

The main functions of the HW4 software include on-line data monitoring and recording by the PC, automatic generation of graphs and data tables, configuration of the HygroClip DI, and adjustment (calibration) of the HygroClip probes connected to the HygroClip DI.

The main features of the HygroClip DI are as follows:

Two probe inputs (HygroClip probes)

Choice of USB or Ethernet interface - wired or wireless

RS-485 port for connecting up to 64 devices to a multi-drop network.

Power for the HygroClip DI is normally provided by an external AC adapter. The RS-485 port also allows powering the HygroClip DI and other devices in a multi-drop network from a single DC power source (user supplied).

2.Models

HygroClip DI-3

USB port

Two probe inputs (HygroClip probes)

RS-485 port (4-pin connector: TX +, TX - , DC in, GND)

Receptacle for AC adapter

© 2007; Rotronic AG

IN-E-HyClip DI-V1_10.doc

Image 3
Contents HygroClip DI Digital Interface IN-E-HyClip DI-V110HygroClip DI digital interface instruction Manual Contents Overview ModelsHygroClip DI-3 HygroClip DI-4-WL Power supplyProbes HygroClip DI-4IN-E-HyClip DI-V110 Connector identification Operation with the HW4 software Pin-out diagramsHW4 Device Manager File Device Manager Menu BarDevice Information Language / Unit System Probe Inputs Digital Interface Compliance with FDA 21CFR Part II and Gamp IN-E-HyClip DI-V110 IN-E-HyClip DI-V110 Baud rate compatibility requirements RS-485 multi-drop network Ethernet local area networkUSB Example HygroClip DI with USB portSpecifications Communications protocol AccessoriesPower supply wiring Electrical Installation GuidelinesLightning protection Hardware and firmware Document releases

HygroClip DI digital interface specifications

The Digi HygroClip DI is an advanced digital interface designed for precise measurement and control of humidity and temperature in various applications. An integral tool for numerous industries, the HygroClip DI is particularly prized in environmental monitoring, HVAC systems, agriculture, and laboratories, where accurate climate data is essential for operational efficiency and sustainability.

One of the standout features of the Digi HygroClip DI is its exceptional measurement accuracy. The device utilizes a high-performance capacitive sensor for humidity measurement and a precision thermistor for temperature. This combination enables the HygroClip DI to deliver consistent and reliable readings, ensuring that users can trust the data used for critical decision-making processes.

The HygroClip DI incorporates innovative digital signal processing technology. This technology enhances the device's stability and response time, significantly improving performance compared to traditional analog sensors. This digital approach also simplifies integration with other systems, making it an ideal choice for applications requiring connectivity to data loggers, control systems, or telemetry networks.

Another significant characteristic of the HygroClip DI is its robust design. Constructed to withstand demanding environments, the device is both water-resistant and resistant to dust, ensuring that it maintains functionality even in challenging conditions. Its design also allows for easy installation, often requiring minimal effort to mount or connect.

Flexibility is another key advantage of the Digi HygroClip DI. It comes with various calibration options and can accommodate different operating ranges, making it versatile for diverse industrial needs. Users can select calibration settings that best match their specific environment, further enhancing the accuracy and efficiency of the device.

Additionally, the HygroClip DI is equipped with intuitive interfaces for easy operation. With user-friendly software options available, users can easily manage settings, monitor conditions, and analyze data trends. This accessibility makes it suitable for both experienced professionals and newcomers to the field.

Overall, the Digi HygroClip DI embodies a perfect blend of technology, accuracy, and user-friendliness. Its features cater to a wide range of measurement needs, making it an invaluable tool for any professional seeking reliable climatic data in their operations. Whether used in industrial settings or research facilities, the HygroClip DI is a testament to innovation in environmental monitoring technology.