Digi HygroClip DI digital interface instruction manual Example HygroClip DI with USB port, Usb

Page 19

IN-E-HyClip DI-V1_10

Rotronic AG

Bassersdorf, Switzerland

Document code

Unit

 

 

 

HygroClip DI digital interface: instruction

Instruction Manual

manual

 

Document Type

Document title

Page 19 of 24

 

 

 

 

 

for RS-485 slaves” is being used, HW4 automatically changes from 00 to 01 the RS-485 address of any master that has slaves attached. As a result, all masters end up with the same RS-485 network address (01). In addition, HW4 automatically changes the RS-485 address of the slaves in each multi-drop from 00 (factory default) to a unique address ranging from 02 to 64. The same address range (02 to 64) is used again when there is more than one multi-drop RS485 network.

Notes:

-We recommend using HW4 to give each HygroClip DI a unique name for easier identification.

-In the situation where an RS-485 network is to be monitored by software other than HW4, you will have to use HW4 to give each device a unique RS-485 address prior to connecting to the network.

Example: HygroClip DI with USB port

The number of HygroClip DI with USB port that can be directly connected to a PC depends on the number of USB ports available on the PC.

PC (HW4)

USB port

DI

USB

Master

Address 1

RS485

RS485: 2 wires

Slave

Address 2

RS485

Slave

Address 64

RS485

IMPORTANT:

Multi-drop: up to 1000m (3300 ft)

-If you change an instrument from master to slave, be sure to change the RS address of the instrument to 00 before establishing the RS-485 connection. Not doing so may prevent HW4 from detecting the instrument. After detecting the instrument, HW4 will automatically change the RS address to 02 or higher. Please note that HW4 may eventually change the address of any master with address 00.

-All devices within a multi-drop should use the same Baud rate. See Baud rate compatibility requirements.

-HygroClip DI-4 and DI-4-WL: please note that the Baud rate of the HygroClip DI and the Baud rate of the internal Ethernet module should be set to the same value

-The RS485 wiring can be done with a twisted pair (2-wire, telephone type) and wall mounted RJ45 receptacles. No shielded cable and no terminator should be required. Be sure to observe the polarity for each connection. Using 4-wire cable, offers the option of powering all devices in the multi-drop from a single DC voltage source.

© 2007; Rotronic AG

IN-E-HyClip DI-V1_10.doc

Image 19
Contents IN-E-HyClip DI-V110 HygroClip DI Digital InterfaceHygroClip DI digital interface instruction Manual Contents Models OverviewHygroClip DI-3 HygroClip DI-4-WL Power supplyProbes HygroClip DI-4IN-E-HyClip DI-V110 Connector identification Operation with the HW4 software Pin-out diagramsHW4 Device Manager File Device Manager Menu BarDevice Information Language / Unit System Probe Inputs Digital Interface Compliance with FDA 21CFR Part II and Gamp IN-E-HyClip DI-V110 IN-E-HyClip DI-V110 Baud rate compatibility requirements RS-485 multi-drop network Ethernet local area networkUSB Example HygroClip DI with USB portSpecifications Communications protocol AccessoriesPower supply wiring Electrical Installation GuidelinesLightning protection Hardware and firmware Document releases

HygroClip DI digital interface specifications

The Digi HygroClip DI is an advanced digital interface designed for precise measurement and control of humidity and temperature in various applications. An integral tool for numerous industries, the HygroClip DI is particularly prized in environmental monitoring, HVAC systems, agriculture, and laboratories, where accurate climate data is essential for operational efficiency and sustainability.

One of the standout features of the Digi HygroClip DI is its exceptional measurement accuracy. The device utilizes a high-performance capacitive sensor for humidity measurement and a precision thermistor for temperature. This combination enables the HygroClip DI to deliver consistent and reliable readings, ensuring that users can trust the data used for critical decision-making processes.

The HygroClip DI incorporates innovative digital signal processing technology. This technology enhances the device's stability and response time, significantly improving performance compared to traditional analog sensors. This digital approach also simplifies integration with other systems, making it an ideal choice for applications requiring connectivity to data loggers, control systems, or telemetry networks.

Another significant characteristic of the HygroClip DI is its robust design. Constructed to withstand demanding environments, the device is both water-resistant and resistant to dust, ensuring that it maintains functionality even in challenging conditions. Its design also allows for easy installation, often requiring minimal effort to mount or connect.

Flexibility is another key advantage of the Digi HygroClip DI. It comes with various calibration options and can accommodate different operating ranges, making it versatile for diverse industrial needs. Users can select calibration settings that best match their specific environment, further enhancing the accuracy and efficiency of the device.

Additionally, the HygroClip DI is equipped with intuitive interfaces for easy operation. With user-friendly software options available, users can easily manage settings, monitor conditions, and analyze data trends. This accessibility makes it suitable for both experienced professionals and newcomers to the field.

Overall, the Digi HygroClip DI embodies a perfect blend of technology, accuracy, and user-friendliness. Its features cater to a wide range of measurement needs, making it an invaluable tool for any professional seeking reliable climatic data in their operations. Whether used in industrial settings or research facilities, the HygroClip DI is a testament to innovation in environmental monitoring technology.