FieldServer FS-8700-123 instruction manual Appendix A. Driver Error Messages

Page 21

FS-8700-123_MetOne Particle Counter by Hach Ultra Analytics Manual Page 21 of 26

Appendix A. Driver Error Messages

 

Message

 

 

 

 

Description

 

 

 

 

 

 

An attempt was made to send a mode control command to

 

 

 

 

 

 

a device. The required mode is derived from the Data

 

 

 

 

 

Array value. This message is printed when the driver

 

MET1:#01 Err. Cant set mode to

 

doesn’t recognize the value found and thus cannot send

 

state=%d for MD=%s

 

 

 

the command.

This can be resolved by having the

 

 

 

 

 

 

upstream device send a valid value to the DA:offset

 

 

 

 

 

 

associated with the reported MD. Refer to the example in

 

 

 

 

 

 

section 4.9 for more information.

 

 

 

 

 

 

The reported Map Descriptor is being used to send a new

 

 

 

 

 

 

Sample Time to the field device but the time found in the

 

MET1:#02 Err. Cant set Sample

 

Data Aarray:offset specified on the Map Descriptor is zero.

 

time to zero. MD=%s

 

 

 

The driver cannot send a sample time of zero. This can be

 

 

 

 

 

 

resolved by by having the upstream/other protocol send a

 

 

 

 

 

 

non-zero value to the DA:Offset.

 

 

 

 

 

 

The reported Map Descriptor is being used to send a new

 

 

 

 

 

 

Hold Time to the field device but the time found in the Data

 

MET1:#03 Err. Cant set Holde

 

Array:offset specified on the Map Descriptor is zero. The

 

time to zero. MD=%s

 

 

 

driver cannot send a sample time of zero. This can be

 

 

 

 

 

 

resolved by by having the upstream/other protocol send a

 

 

 

 

 

 

non-zero value to the DA:Offset.

 

 

 

 

 

 

Message 4a is printed when a response to a poll for the

 

 

 

 

 

 

Sample/Hold time returned a response whose time cannot

 

 

 

 

 

 

be correctly understood by the driver. If this error occurs

 

 

 

 

 

 

rarely it may be a consequence of an occasional

 

 

 

 

 

 

corruption. If it occurs consistently, take a log, ensure this

 

MET1:#04a/b

Err.

Invalid

time

 

event occurs at least once during the log and contact Tech

 

string=%s MD=%s

 

 

 

Support. The FieldServer web site provide tools and

 

 

 

 

 

 

instructions for taking logs.

 

 

 

 

 

 

Message 4b is printed when the time is formatted correctly

 

 

 

 

 

 

but found be an invalid time. The same considerations

 

 

 

 

 

 

apply as with 4a.

 

 

MET1:#05 FYI. MD=%s too short

 

This message is printed once and trhen suppressed.

 

to store

all. Ignored

stat

 

 

 

 

incremented

each

time

this

 

The Map descriptor is not long enough to store all data

 

happens. Play safe - set MD

 

from the record..2

Refer to section 4.11.

 

length=20.

 

 

 

 

 

 

 

 

 

 

 

 

If a response to a read Current Record contains a

 

 

 

 

 

 

checksum error the driver prints this message.

 

MET1:#06 Err. Cant store Current

 

If this error occurs rarely then it may be a consequence of

 

Record. Chksum err.

 

 

 

an occasional corruption. If it occurs consistently, take a

 

 

 

 

log, ensure this event occurs at least once during the log

 

 

 

 

 

 

 

 

 

 

 

 

and contact Tech Support. The FieldServer web site

 

 

 

 

 

 

provide tools and instructions for taking logs

2Update the configuration file, download the modified configuration file and reset the FieldServer for the changes to take effect.

FieldServer Technologies 1991 Tarob Court Milpitas, California 95035 USA Web:www.fieldServer.com Tel: 408 262-2299 Fax: 408-262-2269 Toll_Free: 888-509-1970 email: support@fieldServer.com

Image 21
Contents Applicability & Effectivity MetOne Particle CountersTable of Contents MetOne Particle Counter Serial Description FS-8700-123 MetOne Particle Counter Serial Driver Manual Hardware Connections RS-232 Port Hardware Connection Tips / HintsData Arrays/Descriptors Configuring the FieldServer as a MetOne PC Serial ClientClient Side Connection Descriptions Client Side Node DescriptorsHF CNC Timing Parameters Driver Related Map Descriptor ParametersDatimes Rdbc MET1 Datimes Wrbx MET1 Map Descriptor Example 3 Read the operating modeMap Descriptor Example 4 Start / Stop Counting Map Descriptor Example 5 Change Mode Map Descriptor Example 6 Select Device Map Descriptor Example 7 Read the Current Record Map Descriptor Example 8 Dumping the Current Record DACURRD01LBL ,0 Map Descriptor Example 10 Reading from a SubDevice/Port Map Descriptor Example 11 Selecting a Sub-Device/Port Configuring the FieldServer as a MetOne PC Serial Server Appendix A. Driver Error Messages Page FYI Appendix A.1. Driver Operating and Error StatisticsUINT32 Appendix A.2. Driver Exposed Operating and Error StatisticsPage This page Intentionally Left Blank