ATS HILT 9000 manual Quality Q

Page 7

6

www.atsi-tester.com

The LOOP test can measure a maxi- mum Q of 15. If Q is greater than 15 the tester displays Q>15. To measure the actual value, press the Q button on the keypad to run a separate Q test.

The default frequency at which the LOOP and Q tests run is 50 KHz. It can be changed to any value within the range from 20 KHz to 80 KHz with 1 KHz increment.

R=12.3Q=12.1 L=89.2µH 50KHz

Figure 3

Quality (Q)

ATTENTION

Disconnect the detector from the loop before running this test.

DO NOT switch to a different test while the measure- ments are being generated for the current test. Only after the results of the current test are displayed can a different test be performed.

Connect the test leads to the loop. This test requires several seconds to run during which the tester will measure Q at some arbitrary frequency and calculate Q for a user-selected

frequency. If a user-selected frequency is not entered prior to

pressing Q, the default 50KHz setting

 

 

will be used. Refer to the Set

 

 

Frequency section on page 7 for

Q=12.1

@40.5KHz

full instructions to enter a different

*Q=14.9

@50.0KHz

frequency.

 

Both Q results will be presented on the

 

display as shown in Figure 4. The

Figure 4

calculated value is shown on the

 

second line and indicated with an *.

 

Image 7
Contents Hilt Handheld Inductive Loop Tester Table of Contents Packaging Receiving your ShipmentGeneral Description Detector Measures operating parameters Overview of Hilt 9000 ControlsLoop Function Controls PowerQuality Q Change of Inductance Set FrequencyDetector Button Description Comments Function Controls SummaryOverview of Traffic Loop Detection Systems Battery replacement Commonly Asked Questions When I run a Q test, why are there two results shown?Atsi Repair and CalibrationTechnical and Sales Assistance This page left blank Warranty Repair Limited Warranty