Practical Application Examples
Absolute output power measurements with flatness correction
One benefit of the
Table 6. Absolute output power measurements with flatness correction
8510C Keystrokes | Description |
Set up the source and power meter as illustrated in Figure 2
Set up the measurement
PARAMETER [MENU] | Set up the analyzer for a b2 measurement. |
{USER 2 b2} |
|
Activate the test port power flatness-correction
Perform the flatness correction calibration. Set the test port power level and enable the flatness correction as explained in Table 1.
Connect a thru
Perform a thru calibration with the flatness correction enabled
[CAL] | Perform a thru calibration to eliminate the frequency response errors of the port 2 path in the measurement. Be sure to |
{CAL#…} | include any attenuators and/or adapters which are part of the measurement in the thru calibration. It may be necessary to |
{CALIBRATE:RESPONSE} | swap adapters for the thru connection. Any cal set may be selected to access the response calibration in the calibration |
{THRU} | menu structure |
{DONE RESPONSE} |
|
{CAL SET #} |
|
Connect the DUT
Measure the absolute output power
RESPONSE [MENU] | When the device is reconnected, the gain will be displayed. Enter a magnitude offset equivalent to the test port power level |
{MORE} | (P2). Measure the absolute output power at any point in the measurement span. |
{MAGNITUDE OFFSET} |
|
P2 [x1] |
|
10