Agilent Technologies PN 8510-16 manual Flatness-correction calibration procedure

Page 5

Table 1. Flatness-correction calibration procedure

8510C Keystrokes Description

Set up the power meter (see Table 2)

 

For proper operation, the E4418B must be set up before initiating the flatness-correction routine.

 

See Table 2 for specific instructions.

Verify the power meter's address on the 8510 system

[LOCAL]

When shipped from the factory, the address of the power meter is 13. If this conflicts with another instrument on the system bus,

{POWERMETER}

change the address of the conflicting instrument. The power meter must be set to address 13 in order for the firmware to func-

 

tion properly. Press [SYSTEM/INPUTS], {REMOTE INTERFACE} and {INTERFACE OVERVIEW} on the E4418B to ensure the

 

power meter’s address is set to 13. Verify that the power meter is connected to the system bus.

Set up the analyzer and measurement

Set up the start/stop frequencies and the measurement type (for example, S11, S12).

Adjust the number of analyzer trace points (if needed)

STIMULUS [MENU]

When the flatness calibration is initiated, the analyzer sends the source a list of flatness-correction frequencies equal to the

{NUMBER OF POINTS}

number of trace points set on the analyzer.

{POINTS}

 

Set the source to the maximum leveled power

STIMULUS [MENU]

Set the 8360 for maximum specified power (P1) at the highest frequency in the span; see Table 3. At higher power levels, the

{POWER MENU}

power meter will take less time to settle between measurement points during the calibration process.

{POWER SOURCE 1}

 

P1 [x1]

 

Connect the power sensor to the active test port and begin the flatness calibration

Initiate flatness calibration

{POWER FLATNESS}

Do not cycle the power on the analyzer or the source during the calibration process. The calibration may be aborted at any time by

{CALIBRATE FLATNESS}

pressing any key on the analyzer’s front panel. Aborting the calibration is not recommended, since the command may cause the

 

system to freeze. By pressing {CALIBRATE FLATNESS} the analyzer will remind the user to zero the power meter and connect the

 

power sensor to the source. Press {CALIBRATE FLATNESS} again on the analyzer to initiate the calibration. Once initiated, the

 

analyzer retrieves the power meter measurement data and transfers it to the 8360 where it is processed and stored with the

 

appropriate correction frequency. During the flatness-correction calibration, the analyzer will display MEASURING DATA, xx%

 

DONE, PRESS ANY KEY TO ABORT. The xx% DONE indicates the percentage of the total number of correction points that have

 

been measured; it is not an indication of elapsed measurement time. When the flatness-correction calibration is completed, the

 

analyzer will automatically store the correction table into register 1 of the source. Any new calibration of the source will overwrite

 

the flatness-correction table.

Activate the flatness-correction

{FLATNESS ON}

The source output power will now be unleveled as it attempts to output the test port power level (Power Source 1) plus the flat-

 

ness correction for each measurement point in the frequency span. IF Overload or Source 1 Warning—RF Unleveled may be dis-

 

played on the analyzer until the test power level is reduced.

Set the test port power level

STIMULUS [MENU]

Set the test port power level (P2) equal to or below the maximum allowed test port power (see Table 3) for the highest frequency

{POWER MENU}

in the measurement span. A constant power level equivalent to P2 will now be available at the test port. The flatness-correction

{POWER SOURCE 1}

calibration can be varified using the E4418B to measure the power at certain CW frequencies in the frequency range.

P2 [x1]

 

Perform the measurement calibration

Compensate for systematic errors by performing a measurement calibration (if desired). Although a measurement calibration can be performed before or after the flatness-correction calibration with or without correction enabled, it is recommended that the cal- ibration be performed after the flatness correction is enabled. The calibration is no longer valid if the source power is changed after performing the calibration.

Connect the DUT to the test port

5

Image 5
Contents Uncorrected Power Corrected Power Output Power Agilent PN Controlling Test Port Output Power FlatnessIntroduction 85107B Measurement System System ConfigurationFlatness-Correction Operation Flatness-correction calibration procedure Set up the 437B emulation on the power meter Agilent E4418B power meter setup83620B 83621B 83623B 83623L 83631B 83651B FrequencyCalibration time Power control with flatness correctionOperational Considerations Verifying a flatness-correction calibration When to recalibrateUsing test port 1 calibrations on test port Activate the test port power flatness-correction Absolute output power measurements with flatness correctionConnect the DUT Measure the absolute output power Practical Application ExamplesConnect the amplifier and measure the absolute output power Set up a b2 measurement and connect a thruAdjust the display Increase power to find the 1 dB gain compression pointOnline Assistance Our PromiseYour Advantage