Agilent Technologies E3630A service manual Remove short and proceed to

Page 18

Table A-5. +20V Supply Troubleshooting (Cont’d)

Low output voltage

1.

Measure voltage at the

a.

Measured voltage is less than 0

a. Check for open Q9, Q6, R26, or

(lower than rating)

 

base of Q5.

 

volt.

 

CR5.

 

 

 

b.

Measures voltage is more than 0

b.

Proceed to step (2).

 

 

 

 

volt.

 

 

 

2.

Eliminate current limit

a.

Output voltage increases.

a. Check for U6B defective.

 

 

circuit as a source of

b.

Output voltage remains low.

b. Reconnect lead and proceed to

 

 

trouble by disconnecting

 

 

 

step (3).

 

 

anode of CR11.

 

 

 

 

 

3.

Measure voltage at pin 3

a.

Measured voltage is near

a.

Check for defective U6A.

 

 

of U6.

 

-0.7V.

 

 

 

 

 

b.

Measured voltage is zero volt.

b. Check for CR14 or CR15 shorted.

 

 

 

 

 

c. Check for open R39, shorted R36,

 

 

 

c.

Measured voltage is near

 

or leaky or shorted C7.

 

 

 

 

+0.7V.

 

 

 

 

 

 

 

 

 

 

Table A-6. -20V Supply Troubleshooting

 

 

 

 

 

 

 

 

 

SYMPTOM

 

STEP - ACTION

 

RESPONSE

 

PROBABLE CAUSE

 

 

 

 

 

The +20V supply must operate properly before troubleshooting the -20V supply.

 

 

 

 

 

 

 

High output voltage

1.

Attempt to turn off Q8 by

a.

Output voltage remains high.

a.

Q8 shorted.

(more than 1%

 

shorting emitter-to-base

b.

Output voltage becomes near zero

b.

Remove short and proceed to

greater than +20V

 

of Q8.

 

volt.

 

step (2).

supply in fixed

 

 

 

 

 

 

tracking ratio mode)

2.

Measure voltage at

a.

Measured voltage is more than

a. Check for open CR20 or R14 and

 

 

base of Q7.

 

0 volt.

 

check for defective U8A.

 

 

 

b.

Measured voltage is less than

b.

Check for defective Q7.

 

 

 

 

0 volt.

 

 

 

 

 

 

 

 

Low output voltage

1.

Measure voltage at the

a.

Measured voltage is more than 0

a. Check for open Q8, Q3, and R24.

(more than 1% lower

 

base of Q7.

 

volt.

b.

Proceed to step (2).

than +20V supply in

 

 

b.

Measured voltage is less than

 

 

fixed tracking ratio

 

 

0

volt.

 

 

mode)

 

 

 

 

a. Check for U8B defective.

 

2.

Eliminate current limit

a.

Output voltage increases.

b. Reconnect lead and proceed to

 

 

circuit as a source of

b.

Output voltage remains low.

 

step (3).

 

 

trouble by disconnecting

 

 

 

 

 

 

anode of CR18.

 

 

 

 

 

 

 

 

 

a. Check for defective U8A.

 

3.

Measure voltage at pin 3

a.

Measured voltage is near

 

 

 

 

of U8.

 

+0.7V.

b. Check for shorted CR21 and

 

 

 

b.

Measured voltage is zero volt.

 

CR22.

 

 

 

 

 

c. Check for open R15, shorted R41,

 

 

 

c.

Measured voltage is near

 

or leaky or shorted C6.

 

 

 

 

-0.7V.

 

 

 

 

 

 

 

 

 

A-9

Image 18
Contents Dual Output Power Supply AprilDo not Service or Adjust Alone Safety SummaryBefore Applying Power Safety Symbols Ground the InstrumentTable of Contents General Information Specifications Line FuseInput Power Requirements InstallationInstallation Data Operating InstructionsTracking Ratio Control Turn -ON Checkout ProcedureOperation Overload Protection CircuitsReverse Current Loading Load ConsiderationsPulse Loading Output CapacitanceReverse Current Loading Solution Service Information Principles of OperationMaintenance Type Required Characteristics USE Recommended ModelTest Equipment Required Performance Tests Rated Output, Tracking, Meter Accuracy, and Cur- rent Limit Figure A-4. Current Sampling Resistor ConnectionsLoad Regulation Load Effect Common Mode Current CMI Load Transient Response TimeOpen Fuse Troubleshooting TroubleshootingInitial Troubleshooting Procedure Symptom Check Probable CauseSymptom Step Action Response Probable Cause Step Action Response Next ActionRemove short and proceed to Step Action Response Probable Cause Adjustment and Calibration Current Limit AdjustmentMeter Calibration Replaceable Parts Ordering InformationCode Manufacturer Address 6CDNG#4GRNCEGCDNG2CTVU.KUV 6CDNG#4GRNCEGCDNG2CTVU.KUV%QPVF 45+5614-96% 100%6142156622+052%%106 Certification Seoul 153-023 Korea Quality Manager