Agilent Technologies E3630A service manual Common Mode Current CMI, Load Transient Response Time

Page 15

Peak-to-Peak Measurement

The peak-to-peak measurement is particularly important for applications where noise spikes could be detrimental to a sensitive load, such as logic circuitry.

To measure the ripple and noise of the peak-to-peak value on each output supply output:

a.Connect the test equipment across the output of the +20V supply as shown in Figure A-7, but replace the true rms voltmeter with the oscilloscope.

b.Turn on the supply and push +20V METER switch.

c.Turn up output voltage to the full rated value.

d.Set the oscilloscope to AC mode and bandwidth to 20 MHz.

e.Check that the peak-to-peak noise is less than 1.5 mV.

f.Repeat for the remaining supply outputs.

Common Mode Current (CMI)

Definition : Common mode current is that ac current compo- nent which exists between any or all supply or output lines and chassis ground.

To measure the common mode current:

a.Connect the full load for +6V output.

b.Connect a 100 kW resistor(RS) and a 2200 pF capacitor in parallel between common terminal(COM) and chassis ground.

c.Connect the DVM across RS.

d.Turn on the supply.

e.Record the voltage across RS and convert it to current by dividing this voltage by RS.

f.Check that the current is less than 1 mA.

Load Transient Response Time

Definition : This is the time for the output voltage to return to within a specified band around its voltage following a change from full load to half load or half load to full load.

To measure the load transient response time:

a.Connect the test equipment across the output of the

+20V supply as shown in Figure A-6, but replace the DVM with the oscilloscope. Operate the electronic load in constant current mode.

b.Turn on the supply.

c.Turn up output voltage to the full rated value.

d.Set the electronic load to transient operation mode between one half of supply's full rated value and supply's full rated value at a 1 KHz rate with 50% duty cycle.

e.Set the oscilloscope for ac coupling, internal sync and lock on either the positive or negative load transient.

f.Adjust the oscilloscope to display transients as in Figure A-8.

g.Check that the pulse width (t2-t1) of the transients at 15 mV from the base line is no more than 50 msec as shown.

h.Repeat for the remaining supply outputs.

Figure A-8. Load Transient Response Time Waveform

A-6

Image 15
Contents April Dual Output Power SupplyGround the Instrument Safety SummaryBefore Applying Power Safety Symbols Do not Service or Adjust AloneTable of Contents General Information Line Fuse SpecificationsOperating Instructions InstallationInstallation Data Input Power RequirementsOverload Protection Circuits Turn -ON Checkout ProcedureOperation Tracking Ratio ControlOutput Capacitance Load ConsiderationsPulse Loading Reverse Current LoadingReverse Current Loading Solution Principles of Operation Service InformationMaintenance Type Required Characteristics USE Recommended ModelTest Equipment Required Performance Tests Figure A-4. Current Sampling Resistor Connections Rated Output, Tracking, Meter Accuracy, and Cur- rent LimitLoad Regulation Load Effect Load Transient Response Time Common Mode Current CMISymptom Check Probable Cause TroubleshootingInitial Troubleshooting Procedure Open Fuse TroubleshootingStep Action Response Next Action Symptom Step Action Response Probable CauseRemove short and proceed to Step Action Response Probable Cause Adjustment and Calibration Current Limit AdjustmentMeter Calibration Replaceable Parts Ordering InformationCode Manufacturer Address 6CDNG#4GRNCEGCDNG2CTVU.KUV 6CDNG#4GRNCEGCDNG2CTVU.KUV%QPVF 45+5614-96% 100%6142156622+052%%106 Certification Quality Manager Seoul 153-023 Korea