put is enclosed by a heavy line in Figure 3 of the operating section. If the operating point reaches the diagonal current limit line, a decrease in load resistance moves the operating point down the line, reducing the output voltage and current. Current foldback is controlled by a second operational ampli- fier (current comparison amplifier) in the regulator that moni- tors the dc output current. This current comparison amplifier takes control of the output away from the voltage comparison amplifier when the current reaches the design limit. Removing the overload restores constant voltage operation auto- matically.
The 0 to
When output voltage exceeds the reference level, the SCR control circuit fires two SCRs that cause the input capacitor to be charged to the voltage which is necessary for full output of the supply. When the two SCRs are not fired, the bridge diode CR26 conducts and half the voltage is applied to series pass transistor Q9.
The 0 to
The reference and bias supply powers the operation amplifi- ers and provides reference and bias voltages for the output regulators. The display power circuit provides voltage which is used by the A/D converter and display.
The
Three meter
Diode CR2, CR3, and CR4 are connected across the output terminals in reverse polarity. They protect the output elec- trolytic capacitor and the series regulator transistors from a reverse voltage applied across the output terminals.
MAINTENANCE
INTRODUCTION
This section provides performance test, troubleshooting infor- mation, and adjustment and calibration procedures. The fol- lowing operation verification tests comprise a short procedure to verify that the power supply is performing properly, without testing all specified parameters.
If a fault is detected in the power supply while making the performance check or during normal operation, proceed to the troubleshooting procedures. After troubleshooting, per- form any necessary adjustments and calibrations. Before returning the power supply to normal operation, repeat the performance check to ensure that the fault has been properly corrected and that no other faults exist.
Test Equipment Required
The following Table
Table
TYPE | REQUIRED CHARACTERISTICS | USE | RECOMMENDED |
|
|
| MODEL |
|
|
|
|
Oscilloscope | Sensitivity : 100 mV | Display transient response and ripple | Agilent 54600A |
| Bandwidth : 20 MHz/100 MHz | and noise waveforms. |
|
|
|
|
|
RMS Voltmeter | True rms, 20 MHz bandwidth | Measure rms ripple and noise |
|
| Sensitivity : 1 mV | voltage. |
|
| Accuracy : 5% |
|
|
|
|
|
|
Multimeter | Resolution : 100 nV | Measure dc voltages. | Agilent 34401A |
| Accuracy : 0.0035% |
|
|
|
|
|
|
Electronic Load | Voltage Range : 240 Vdc | Measure load and line regulation. | Agilent 6063A |
| Current Range : 10 Adc |
|
|
| Open and short switches |
|
|
| Transient on/off |
|
|
|
|
|
|
Resistive Loads (RL) | 40 W 20 W, 2.4 W 20 W | Measure ripple and noise. |
|
|
|
|
|