Agilent Technologies E3630A service manual Line Fuse, Specifications

Page 5

mode. In this mode the voltage of the negative output can be set lower than that of the positive output. The tracking ratio control allows the negative supply's output to be set to any value between a maximum that is within ±5% of the positive supply's output and a minimum that is less than 0.5 volts. Once a ratio is established by the tracking ratio control, the ratio of the positive output voltage to the negative output voltage remains constant as the ±20V voltage control varies the 0 to +20V output over its range.

The front panel also contains a line switch, three overload indicators for +6V output, +20V output, and -20V output, a voltmeter and an ammeter, and three push-button meter switches. The push buttons select one of the supply's three outputs for display. The voltmeter and ammeter always moni- tor any one supply simultaneously. In addition to the standard 115 Vac ± 10% 47 to 63 Hz input, two other line voltage options

are available for nominal inputs of 100 and 230 Vac. The sup- ply is furnished with a detachable 3-wire grounding type line cord. The ac line fuse is in an extractor type fuseholder on the rear heat sink.

LINE FUSE

The line fuse is located by the ac line receptacle. Check the rating of the line fuse and replace it with the correct fuse if necessary as indicated below. These are slow-blow fuses.

Line Voltage

Fuse

Agilent Part No.

100/115 Vac

1.6 A

2110-0918

230 Vac

1.0 A

2110-0599

SPECIFICATIONS

Table 1 lists detailed specifications for the power supply.

Table 1. Specifications

AC INPUT

Standard:

115 Vac ± 10%, 47-63 Hz, 115 VA, 84 W

OE9:

100 Vac ± 10%, 47-63 Hz, 115 VA, 84 W

OE3:

230 Vac ± 10%, 47-63 Hz, 115 VA, 84 W

DC OUTPUT and OVERLOAD PROTECTION

0 to ±20 V Outputs: Maximum rated output current is 0.5 A. Short circuit output current is 0.55 A ± 5% and a fixed cur- rent limit circuit limits the output of each supply to this maxi- mum at any output voltage setting. Unbalanced loads within current rating are permitted.

OPERATING TEMPERATURE RANGE

0 to 40 oC for full rated output. At higher temperatures, out-

put current is derated linearly to 50% at 55 oC maximum temperature.

TEMPERATURE COEFFICIENT

All Outputs: Less than 0.02% plus 1 mV voltage change

per oC over the operating range from 0 to 40 oC after 30 minutes warm-up.

STABILITY (OUTPUT DRIFT)

0 to +6 V Output: Maximum rated output current is 2.5 A at 6 V. The maximum available output current decreases with the output voltage setting. A current foldback circuit limits the output to 2.75 A ± 5% at 6 volts and, with decreasing voltage, reduces the current limit linearly to 1 A ± 15% at zero volts (short circuited).

TRACKING ACCURACY

The +20 V and -20 V outputs track within 1% with the TRACKING ratio control in the Fixed position. (In variable tracking ratio mode, the negative tracking output can be adjusted from less than 0.5 V to within ±5% of the setting of the positive output.)

LOAD REGULATION

All Outputs: Less than 0.01% plus 2 mV for a full load to no load change in output current.

LINE REGULATION

All Outputs: Less than 0.01% plus 2 mV for any line volt- age change within rating.

PARD (Ripple and Noise)

All Outputs: Less than 0.35 mV rms/1.5 mV p-p (20 Hz-20 MHz).

Common Mode Current (CMI): Less than 1 mA for all

outputs (20 Hz-20 MHz).

All Outputs: Less than 0.1% plus 5 mV (dc to 20 Hz) dur- ing 8 hours at constant line, load and ambient after an initial warm-up time of 30 minutes.

LOAD TRANSIENT RESPONSE TIME

All Outputs: Less than 50 msec for output recovery to within 15 mV of nominal output voltage following a load change from full load to half load, or vice versa.

OUTPUT VOLTAGE OVERSHOOT

All Outputs: During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V if the out- put control is set for less than 1 V. If the con- trol is set for 1 V or higher, there is no overshoot.

METER ACCURACY: ±(0.5% of output + 2 counts) at

25oC±5oC

METER RESOLUTION

All Output: Voltage 10 mV Current 10 mA

DIMENSIONS

212.3mmW x 88.1 mmH x 269.2 mmD

(8.354 inW x 3.469 inH x 10.591 inD)

WEIGHT: 3.8 kg(8.4 lbs) net, 5.1 kg(11.3 lbs) shipping

1-5

Image 5
Contents April Dual Output Power SupplyBefore Applying Power Safety Symbols Safety SummaryDo not Service or Adjust Alone Ground the InstrumentTable of Contents General Information Line Fuse SpecificationsInstallation Data InstallationInput Power Requirements Operating InstructionsOperation Turn -ON Checkout ProcedureTracking Ratio Control Overload Protection CircuitsPulse Loading Load ConsiderationsReverse Current Loading Output CapacitanceReverse Current Loading Solution Principles of Operation Service InformationTest Equipment Required MaintenanceType Required Characteristics USE Recommended Model Performance Tests Figure A-4. Current Sampling Resistor Connections Rated Output, Tracking, Meter Accuracy, and Cur- rent LimitLoad Regulation Load Effect Load Transient Response Time Common Mode Current CMIInitial Troubleshooting Procedure TroubleshootingOpen Fuse Troubleshooting Symptom Check Probable CauseStep Action Response Next Action Symptom Step Action Response Probable CauseRemove short and proceed to Step Action Response Probable Cause Meter Calibration Adjustment and CalibrationCurrent Limit Adjustment Code Manufacturer Address Replaceable PartsOrdering Information 6CDNG#4GRNCEGCDNG2CTVU.KUV 6CDNG#4GRNCEGCDNG2CTVU.KUV%QPVF 45+5614-96% 100%6142156622+052%%106 Certification Quality Manager Seoul 153-023 Korea