Agilent Technologies 6011A, 6010a AC Dropout Buffer Circuit, Multiple Supply System Shutdown, 105

Page 105

AC Dropout Buffer Circuit

This circuit couples, inverts and isolates the DROPOUT signal (received from the A2 Control Board) of status

output terminal J3-19 ( DROPOUT ). The dropout signal indicates loss of primary power, momentary AC dropout, or "brownout" conditions where the AC line voltage drops below approximately 70% normal. The following paragraph provides a brief description of the AC Dropout Buffer circuit. Refer to the Schematic Diagram

The AC Dropout Buffer Circuit receives a DROPOUT signal from the A2 Control Board. This causes the bias voltage supplied to the Dropout Buffer U14A to be pulled down through diode CR4 thus, turning U14A off. This in turn will cause opto-isolator U3 to turn off. Since external pull up resistors are used, terminal J3-19 (DROPOUT) will go high and remain high until the dropout signal from the A2 Control Board is removed.

Multiple Supply System Shutdown

When using more than one 002 Option equipped power supply in a system, it may be desirable to implement a system shutdown. In this configuration, an OVP trip or remote shutdown of a single unit will cause all of the supplies to shut down.

Figure A-9. System Shutdown using Controller Power Supply

Figure A-9 shows one method of system shutdown. The advantages of this method are that one common is used for all status and control lines (useful for controller-operated systems), and the capability of system reset. As shown in

Figure A-9, one supply's OVERVOLTAGE line is connected to the next supply's REMOTE TRIP line, and so on in a continuous chain.

Note

+5V REG/POWER SUPPLY common from Supply 1 can be used instead of the bias voltage from

 

the controller. However, because of current limits of the + 5V REG, no more than four units can

 

be connected together in this configuration. To prevent ground loops, do not parallel connect + 5V

 

REG from more than one supply.

The note on page 101 tells how to determine if a shutdown was initiated through the remote trip line or by a supply's OVP. This allows the controller to determine which supply initiated the shutdown. Following a multiple supply shutdown, each unit can be reset individually or all the REMOTE RESET lines can be tied together for a system reset.

105

Image 105
Contents Autoranging DC Power Supply Agilent Models 6010A, 6011A Certification Safety Summary Safety Symbol Definitions Table of Contents Replaceable Parts Circuit Diagrams TroubleshootingSafety Considerations ScopeItem Description Manual RevisionsIntroduction Calibration ProcedureTest Equipment Required Operation Verification TestsTest Equipment Required Type Required Characteristics USE Recommended Model4KVA Calibration Procedure Initial SetupMON Display Settings VoutIout Common Mode Setup Performance Tests Measurement TechniquesCR4 Current-Monitoring Resistor Setup Constant Voltage CV Tests Basic Test SetupPage RMS Measurement Test Setup, CV Pard Test Peak-To-Peak Measurement Test Setup, CV Pard Test 6010A 6011A Not Applicable6012B 6015A 6010A 6011A 6012B 6015A Load Transient Recovery WaveformConstant Current CC Tests Page CC Pard Test Setup Initial Troubleshooting Procedures TroubleshootingVdc Control Board Test Connector, A2J7Electrostatic Protection Repair and ReplacementA4 FET Board Removal A2 Control Board RemovalA5 Diode Board Removal A3 Front Panel Board RemovalA1 Main Board Removal Overall Troubleshooting Procedure A1 Designator Wire color A3S1 Position Rear ViewA3 Front Panel Assembly Rear View Main Troubleshooting Setup Using the TablesMain Troubleshooting Setup Modified Mains Cord Set For Troubleshooting Front Panel Troubleshooting Troubleshooting No-Output FailuresA2J7-26 A2J7-25 Troubleshooting Bias Supplies A3 Front Panel Board Failure SymptomsPerformance Failure Symptoms Node + N0DE Node +Troubleshooting AC-Turn-on Circuits Power Section BlocksAC Fault Troubleshooting PWM & ClockRelay Enable + OUTTroubleshooting Down Programmer Troubleshooting DC-To-DC ConverterPWM-ON PWM-OFFWaveforms Troubleshooting CC Circuit Troubleshooting CV CircuitON/OFF OFFSET Voltage Setup Measurement Troubleshooting OVP CircuitPage Autoranging Power OverviewSystem Description Regulation & Control Subsystem A and 6015A Simplified Schematic A and 6012B Simplified Schematic Quick Reference Guide to Major Circuits Major Function OperationInput from Output to DP PWM PWM DPProtection Subsystem Input Power SubsystemDC Power Conversion Subsystem Output SubsystemFront Panel Board Page Simplified Front Panel Schematic Reference Designators Replaceable PartsDescription Abbreviations Ordering InformationMain Board Assembly CR1 CB1CR2 CR5A1 Mechanical VR16010A C37 6010A, 6011A, 6015A Not Used Not Used Not Used Page Not Used VR2 VR5VR6 TB2 TB1Not Used Agilent Model VR1 VR2Drvr TTL NOR Dual TS1CR3 CR6FET N-CHAN A5 Mechanical A6 Mechanical Chassis Electrical Not Used Schematic Diagram Notes Component Location and Circuit DiagramsIndicate number of paths represented by the line Top View, Top Covers Removed Main Board A1 and Filter Board A6 Component Location Control Board A2 Component Location Front Panel Board A3 Component Location FET Board A4 Component Location Diode Board A5 Component Location Page Page Page Page Specifications General InformationOption 002 Hardware Input Compliance Voltage ± Current Programming Enable Table A-1. Specifications, Option Remote ProgrammingStatus Indicators AccuracyPower-on Preset Remote Trip and Remote Reset TimingMaximum Output Voltage logic high + On State logic lowBias Supplies DC Output Ratings 25C ± Table A-1. Specifications, Option Pulse TimingShort Circuit Output Current Pard TypicalConnector Assembly Procedure InstallationOperation Figure A-1. Mating Connector AssemblyLocal/Remote Programming Resistance Voltage or Current Figure A-4. Calculating Value of Series Dropping Resistor Remote Resistance ProgrammingFigure A-5. Remote Resistance Programming 101 Remote MonitoringStatus Indicators Remote Control102 Remote Reset Overvoltage103 104 Power-On PresetAC Dropout Buffer Circuit Multiple Supply System Shutdown105 Maintenance Bias Supplies106 Troubleshooting Resistance and Voltage Programming Troubleshooting Current Programming107 108 Figure A-11. Troubleshooting Current Programming of CV Mode109 Figure A-13. Troubleshooting Status Indicators110 111 Table A-3. Replacement Parts112 113 VR9Logic Symbols and Definitions Indicator and Qualifier Symbols114 115 Schematic Diagram NotesFigure A-15. Option 002 Board, Component Location 116 Schematic NotesFigure A-16. Option 002 Board, Schematic Diagram 117 Page Model 6010A Change Model 6011A Change119 Model 6012B Change Model 6015A Change120 121 Delete122 123 124 125 DS5126
Related manuals
Manual 48 pages 27.24 Kb

6015A, 6012B, 6011A, 6010a specifications

Agilent Technologies, a leader in the field of measurement and analysis, offers a suite of instruments within its 6010 and 6011 series, specifically the 6010A, 6011A, 6012B, and 6015A models. These devices are designed to meet the needs of various industries, including healthcare, environmental monitoring, and materials testing.

The Agilent 6010A is a high-performance spectrometer known for its precision and versatility. It utilizes advanced optical technologies to provide exceptional wavelength accuracy and resolution. This model is particularly useful in laboratories where reliable data is critical, offering a wide spectral range and effective noise reduction features. Its user-friendly interface simplifies complex analyses, making it suitable for both seasoned professionals and newcomers.

Following closely, the Agilent 6011A is recognized for its robust capabilities in laboratory environments. This device incorporates advanced signal processing techniques, enabling high-throughput measurements without compromising on quality. The 6011A is ideal for real-time monitoring applications, ensuring that users can make informed decisions based on accurate, timely data. Its comprehensive software suite is designed to enhance data analysis, allowing for seamless integration with existing laboratory workflows.

The 6012B variant enhances the functionality further by introducing additional features tailored for specific applications. With a focus on flexibility, the 6012B supports multiple measurement modes, including direct and differential detection. This model excels in complex measurements, allowing for greater analytical depth and insights. The built-in calibration options ensure consistent performance, making it a reliable choice for various research and development tasks.

Lastly, the Agilent 6015A model stands out with its leading-edge technology, designed for the most demanding applications. It boasts enhanced sensitivity and an improved dynamic range, making it perfect for trace analysis in challenging environmental samples. The 6015A’s advanced reporting tools provide detailed analytics, helping scientists and researchers interpret results efficiently. Its compact design also makes it suitable for laboratory spaces with limited room, without sacrificing performance.

Together, these models showcase Agilent Technologies' commitment to delivering high-quality, innovative solutions that empower users to achieve their analytical goals effectively and efficiently. Whether in a research, clinical, or industrial setting, the 6010A, 6011A, 6012B, and 6015A continue to set standards in precision instrumentation.