Agilent Technologies 6015A Quick Reference Guide to Major Circuits Major Function, Operation

Page 51

Table 4-1. Quick Reference Guide to Major Circuits

Circuit

Major Function

Dependent Circuits

 

Operation

Input from

Output to

Bias Power

Provides Bias and

Mains

Control Circuits

Mains voltage at BVS input is

Supply (BPS)

Reference Voltage.

 

 

converted to lower voltage levels to

 

 

 

 

provide the internal operating voltages

 

 

 

 

for the various circuits.

Bias Voltage

Delays the unit's

BVS

Delay Circuit,

Holds all circuits reset until all internal

Detector (BVD)

operation at power-

 

OVP

voltages are at acceptable levels.

 

on.

 

 

 

Timed Delay

Enables power

BVD: DOD

PWM; Relay

Waits for 3 seconds after power-on and

Circuit (TDC)

circuits.

 

 

then shuts out inrush current limiting

 

 

 

 

resistor. The circuit is triggered by the

 

 

 

 

BVD when the + VDC is stable.

Power Limit

Determines

BVS; Ramp

PWM

Compares VIP RAMP with VREF and

Comparator (PLC)

maximum primary

 

 

produces a signal to inhibit the PWM

 

current.

 

 

when V IP RAMP > V REF.

 

 

 

 

 

Control Voltage

Regulates the

V IP RAMP Control

PWM

Compares VIP RAMP with VCP and

Comparator

operation of the

Port Voltage (VCP)

 

produces a signal to inhibit the PWM

(CVC)

PWM.

 

 

when VIP RAMP > VCP.

Constant Voltage

Produces CV

Outer Voltage

CVC,

Monitors OVS signals from which

Circuit (CV)

Control Voltage.

Sense (OVS)

Display

VMON is derived. Combines OVS and

 

 

Innerloop Voltage

Circuits

IVS to give CV Control Voltage.

 

 

Sense (IVS) CV

 

 

 

 

Program Voltage

 

 

Constant Current

Produces CC

Outer Current

CVC; Display

Monitors OCS signals from which

Circuit (CC)

Control Voltage.

Sense (OCS). CC

Circuits

l-MON is derived. Combines OCS and;

 

 

Program Voltage

 

differentiated IVS to give the CC

 

 

 

 

control voltage.

Pulse Width

Switches FETs.

Master Enable;

FETs

Switching action achieved at 20KHz

Modulator (PWM)

 

PLC, CVC

 

rate with on-pulse activated by 20KHz

 

 

 

 

clock and off-pulse by CVC, PLC,

 

 

 

 

20KHz clock or shutdown circuits.

Primary Current

Generates IP Ramp

FETs

CVC; PLC

Senses Ip current build-up while FETs

Monitor

Voltage.

 

 

are on.

Transformer

 

 

 

 

 

 

 

 

 

Power

Stores and transfers

FETs

Output Rectifier

When FETs are on, the primary

Transformer

output power.

 

 

windings of the transformer store

 

 

 

 

energy until the FETs are switched off

 

 

 

 

when the energy is transferred to the

 

 

 

 

secondary for output circuits.

A4 Q1,2,3,4

Control gating of

PWM

Sensing

FETs open and close in response to

 

current in power,

 

Transformer

pulses from the PWM. The length of its

 

and Sense

 

 

on/off time depends on the duration of

 

Transformers.

 

 

the PWM on or off pulse.

Down

Rapidly lowers

CV Circuit, OVP,

Output Rail

Output filter capacitor are rapidly

Programmer (DP)

output voltage.

DOD

 

discharged at varying ampere rates

 

 

 

 

depending on output voltage. Circuit

 

 

 

 

activated under condition of ac power

 

 

 

 

loss, shut down or low voltage.

51

Image 51
Contents Autoranging DC Power Supply Agilent Models 6010A, 6011A Certification Safety Summary Safety Symbol Definitions Table of Contents Replaceable Parts Scope TroubleshootingCircuit Diagrams Safety ConsiderationsItem Description Manual RevisionsOperation Verification Tests Calibration ProcedureIntroduction Test Equipment RequiredTest Equipment Required Type Required Characteristics USE Recommended Model4KVA Calibration Procedure Initial SetupMON Display Settings VoutIout Common Mode Setup Performance Tests Measurement TechniquesCR4 Current-Monitoring Resistor Setup Constant Voltage CV Tests Basic Test SetupPage RMS Measurement Test Setup, CV Pard Test Peak-To-Peak Measurement Test Setup, CV Pard Test 6010A 6011A Not Applicable6012B 6015A 6010A 6011A 6012B 6015A Load Transient Recovery WaveformConstant Current CC Tests Page CC Pard Test Setup Initial Troubleshooting Procedures TroubleshootingVdc Control Board Test Connector, A2J7Electrostatic Protection Repair and ReplacementA4 FET Board Removal A2 Control Board RemovalA5 Diode Board Removal A3 Front Panel Board RemovalA1 Main Board Removal Overall Troubleshooting Procedure A1 Designator Wire color A3S1 Position Rear ViewA3 Front Panel Assembly Rear View Main Troubleshooting Setup Using the TablesMain Troubleshooting Setup Modified Mains Cord Set For Troubleshooting Front Panel Troubleshooting Troubleshooting No-Output FailuresA2J7-26 A2J7-25 Troubleshooting Bias Supplies A3 Front Panel Board Failure SymptomsPerformance Failure Symptoms Node + N0DE Node +Troubleshooting AC-Turn-on Circuits Power Section Blocks+ OUT Troubleshooting PWM & ClockAC Fault Relay EnablePWM-OFF Troubleshooting DC-To-DC ConverterTroubleshooting Down Programmer PWM-ONWaveforms OFF Troubleshooting CV CircuitTroubleshooting CC Circuit ON/OFFSET Voltage Setup Measurement Troubleshooting OVP CircuitPage Autoranging Power OverviewSystem Description Regulation & Control Subsystem A and 6015A Simplified Schematic A and 6012B Simplified Schematic Quick Reference Guide to Major Circuits Major Function OperationInput from Output to DP PWM PWM DPProtection Subsystem Input Power SubsystemDC Power Conversion Subsystem Output SubsystemFront Panel Board Page Simplified Front Panel Schematic Reference Designators Replaceable PartsDescription Abbreviations Ordering InformationMain Board Assembly CR5 CB1CR1 CR2A1 Mechanical VR16010A C37 6010A, 6011A, 6015A Not Used Not Used Not Used Page Not Used VR2 VR5VR6 TB2 TB1Not Used Agilent Model VR1 VR2Drvr TTL NOR Dual TS1CR3 CR6FET N-CHAN A5 Mechanical A6 Mechanical Chassis Electrical Not Used Schematic Diagram Notes Component Location and Circuit DiagramsIndicate number of paths represented by the line Top View, Top Covers Removed Main Board A1 and Filter Board A6 Component Location Control Board A2 Component Location Front Panel Board A3 Component Location FET Board A4 Component Location Diode Board A5 Component Location Page Page Page Page Specifications General InformationOption 002 Hardware Accuracy Table A-1. Specifications, Option Remote ProgrammingInput Compliance Voltage ± Current Programming Enable Status IndicatorsOn State logic low Remote Trip and Remote Reset TimingPower-on Preset Maximum Output Voltage logic high +Pard Typical Table A-1. Specifications, Option Pulse TimingBias Supplies DC Output Ratings 25C ± Short Circuit Output CurrentConnector Assembly Procedure InstallationOperation Figure A-1. Mating Connector AssemblyLocal/Remote Programming Resistance Voltage or Current Figure A-4. Calculating Value of Series Dropping Resistor Remote Resistance ProgrammingFigure A-5. Remote Resistance Programming 101 Remote MonitoringStatus Indicators Remote Control102 Remote Reset Overvoltage103 104 Power-On PresetAC Dropout Buffer Circuit Multiple Supply System Shutdown105 Maintenance Bias Supplies106 Troubleshooting Resistance and Voltage Programming Troubleshooting Current Programming107 108 Figure A-11. Troubleshooting Current Programming of CV Mode109 Figure A-13. Troubleshooting Status Indicators110 111 Table A-3. Replacement Parts112 113 VR9Logic Symbols and Definitions Indicator and Qualifier Symbols114 115 Schematic Diagram NotesFigure A-15. Option 002 Board, Component Location 116 Schematic NotesFigure A-16. Option 002 Board, Schematic Diagram 117 Page Model 6010A Change Model 6011A Change119 Model 6012B Change Model 6015A Change120 121 Delete122 123 124 125 DS5126
Related manuals
Manual 48 pages 27.24 Kb

6015A, 6012B, 6011A, 6010a specifications

Agilent Technologies, a leader in the field of measurement and analysis, offers a suite of instruments within its 6010 and 6011 series, specifically the 6010A, 6011A, 6012B, and 6015A models. These devices are designed to meet the needs of various industries, including healthcare, environmental monitoring, and materials testing.

The Agilent 6010A is a high-performance spectrometer known for its precision and versatility. It utilizes advanced optical technologies to provide exceptional wavelength accuracy and resolution. This model is particularly useful in laboratories where reliable data is critical, offering a wide spectral range and effective noise reduction features. Its user-friendly interface simplifies complex analyses, making it suitable for both seasoned professionals and newcomers.

Following closely, the Agilent 6011A is recognized for its robust capabilities in laboratory environments. This device incorporates advanced signal processing techniques, enabling high-throughput measurements without compromising on quality. The 6011A is ideal for real-time monitoring applications, ensuring that users can make informed decisions based on accurate, timely data. Its comprehensive software suite is designed to enhance data analysis, allowing for seamless integration with existing laboratory workflows.

The 6012B variant enhances the functionality further by introducing additional features tailored for specific applications. With a focus on flexibility, the 6012B supports multiple measurement modes, including direct and differential detection. This model excels in complex measurements, allowing for greater analytical depth and insights. The built-in calibration options ensure consistent performance, making it a reliable choice for various research and development tasks.

Lastly, the Agilent 6015A model stands out with its leading-edge technology, designed for the most demanding applications. It boasts enhanced sensitivity and an improved dynamic range, making it perfect for trace analysis in challenging environmental samples. The 6015A’s advanced reporting tools provide detailed analytics, helping scientists and researchers interpret results efficiently. Its compact design also makes it suitable for laboratory spaces with limited room, without sacrificing performance.

Together, these models showcase Agilent Technologies' commitment to delivering high-quality, innovative solutions that empower users to achieve their analytical goals effectively and efficiently. Whether in a research, clinical, or industrial setting, the 6010A, 6011A, 6012B, and 6015A continue to set standards in precision instrumentation.