Agilent Technologies 6015A, 6010a, 6011A, 6012B service manual Initial Troubleshooting Procedures

Page 27

3

Troubleshooting

Maintenance described herein is performed with power supplied to the instrument, and protective covers removed. Such maintenance should be performed only by service-trained personnel who are aware of the hazards involved (for example, fire and electrical shock). Where maintenance can be performed without power applied, the power should be removed.

Introduction

Before attempting to troubleshoot this instrument, ensure that the fault is with the instrument itself and not with an associated circuit. The performance test enables this to be determined without having to remove the covers from the supply.

The most important aspect of troubleshooting is the formulation of a logical approach to locating the source of trouble. A good understanding of the principles of operation is particularly helpful, and it is recommended that Chapter 4 of this manual be reviewed before attempting to troubleshoot the unit. Often the user will then be able to isolate a problem simply by using the operating controls and indicators. Once the principles of operation are understood, refer to the following paragraphs.

Table 2-1 lists the test equipment for troubleshooting. Chapter 6 contains schematic diagrams and information concerning the voltage levels and waveforms at many of the important test points. Most of the test points used for troubleshooting the supply are located on the control board test "fingers", which are accessible close to the top of the board. See Table 3-1.

If a component is found to be defective, replace it and re-conduct the performance test. When a component is replaced, refer to Calibration Procedure (Chapter 2). It may be necessary to perform one or more of the adjustment procedures after a component is replaced.

Initial Troubleshooting Procedures

If a problem occurs, follow the steps below in sequence:

a.Check that input power is available, and check the power cord and rear-panel circuit breaker.

b.Check that the settings of mode switch A2S1 are correct for the desired mode of operation. (See Operating Manual).

c.Check that all connections to the power supply are secure and that circuits between the supply and external devices are not interrupted.

d.If the power supply fails turn-on self-test or gives any other indication of malfunction, remove the unit from the operating system before proceeding with further testing.

Some circuits on the power mesh are connected directly to the ac power line. Exercise extreme caution when working on energized circuits. Energize the supply through an isolation transformer to avoid shorting ac energized circuits through the test instrument's input leads. The isolation transformer must have a power rating of at least 4KVA. During work on energized circuits, the safest practice is to disconnect power, make or change the test connections, and then re-apply power.

Make certain that the supply's ground terminal () is securely connected to an earth ground before applying power. Failure to do so will cause a potential shock hazard that could result in personal injury.

27

Image 27
Contents Autoranging DC Power Supply Agilent Models 6010A, 6011A Certification Safety Summary Safety Symbol Definitions Table of Contents Replaceable Parts Scope TroubleshootingCircuit Diagrams Safety ConsiderationsItem Description Manual RevisionsOperation Verification Tests Calibration ProcedureIntroduction Test Equipment RequiredTest Equipment Required Type Required Characteristics USE Recommended Model4KVA Calibration Procedure Initial SetupMON Display Settings VoutIout Common Mode Setup Performance Tests Measurement TechniquesCR4 Current-Monitoring Resistor Setup Constant Voltage CV Tests Basic Test SetupPage RMS Measurement Test Setup, CV Pard Test Peak-To-Peak Measurement Test Setup, CV Pard Test 6010A 6011A Not Applicable6012B 6015A 6010A 6011A 6012B 6015A Load Transient Recovery WaveformConstant Current CC Tests Page CC Pard Test Setup Initial Troubleshooting Procedures TroubleshootingVdc Control Board Test Connector, A2J7Electrostatic Protection Repair and ReplacementA4 FET Board Removal A2 Control Board RemovalA5 Diode Board Removal A3 Front Panel Board RemovalA1 Main Board Removal Overall Troubleshooting Procedure A1 Designator Wire color A3S1 Position Rear ViewA3 Front Panel Assembly Rear View Main Troubleshooting Setup Using the TablesMain Troubleshooting Setup Modified Mains Cord Set For Troubleshooting Front Panel Troubleshooting Troubleshooting No-Output FailuresA2J7-26 A2J7-25 Troubleshooting Bias Supplies A3 Front Panel Board Failure SymptomsPerformance Failure Symptoms Node + N0DE Node +Troubleshooting AC-Turn-on Circuits Power Section Blocks+ OUT Troubleshooting PWM & ClockAC Fault Relay EnablePWM-OFF Troubleshooting DC-To-DC ConverterTroubleshooting Down Programmer PWM-ONWaveforms OFF Troubleshooting CV CircuitTroubleshooting CC Circuit ON/OFFSET Voltage Setup Measurement Troubleshooting OVP CircuitPage Autoranging Power OverviewSystem Description Regulation & Control Subsystem A and 6015A Simplified Schematic A and 6012B Simplified Schematic Quick Reference Guide to Major Circuits Major Function OperationInput from Output to DP PWM PWM DPProtection Subsystem Input Power SubsystemDC Power Conversion Subsystem Output SubsystemFront Panel Board Page Simplified Front Panel Schematic Reference Designators Replaceable PartsDescription Abbreviations Ordering InformationMain Board Assembly CR5 CB1CR1 CR2A1 Mechanical VR16010A C37 6010A, 6011A, 6015A Not Used Not Used Not Used Page Not Used VR2 VR5VR6 TB2 TB1Not Used Agilent Model VR1 VR2Drvr TTL NOR Dual TS1CR3 CR6FET N-CHAN A5 Mechanical A6 Mechanical Chassis Electrical Not Used Schematic Diagram Notes Component Location and Circuit DiagramsIndicate number of paths represented by the line Top View, Top Covers Removed Main Board A1 and Filter Board A6 Component Location Control Board A2 Component Location Front Panel Board A3 Component Location FET Board A4 Component Location Diode Board A5 Component Location Page Page Page Page Specifications General InformationOption 002 Hardware Accuracy Table A-1. Specifications, Option Remote ProgrammingInput Compliance Voltage ± Current Programming Enable Status IndicatorsOn State logic low Remote Trip and Remote Reset TimingPower-on Preset Maximum Output Voltage logic high +Pard Typical Table A-1. Specifications, Option Pulse TimingBias Supplies DC Output Ratings 25C ± Short Circuit Output CurrentConnector Assembly Procedure InstallationOperation Figure A-1. Mating Connector AssemblyLocal/Remote Programming Resistance Voltage or Current Figure A-4. Calculating Value of Series Dropping Resistor Remote Resistance ProgrammingFigure A-5. Remote Resistance Programming 101 Remote MonitoringStatus Indicators Remote Control102 Remote Reset Overvoltage103 104 Power-On PresetAC Dropout Buffer Circuit Multiple Supply System Shutdown105 Maintenance Bias Supplies106 Troubleshooting Resistance and Voltage Programming Troubleshooting Current Programming107 108 Figure A-11. Troubleshooting Current Programming of CV Mode109 Figure A-13. Troubleshooting Status Indicators110 111 Table A-3. Replacement Parts112 113 VR9Logic Symbols and Definitions Indicator and Qualifier Symbols114 115 Schematic Diagram NotesFigure A-15. Option 002 Board, Component Location 116 Schematic NotesFigure A-16. Option 002 Board, Schematic Diagram 117 Page Model 6010A Change Model 6011A Change119 Model 6012B Change Model 6015A Change120 121 Delete122 123 124 125 DS5126
Related manuals
Manual 48 pages 27.24 Kb

6015A, 6012B, 6011A, 6010a specifications

Agilent Technologies, a leader in the field of measurement and analysis, offers a suite of instruments within its 6010 and 6011 series, specifically the 6010A, 6011A, 6012B, and 6015A models. These devices are designed to meet the needs of various industries, including healthcare, environmental monitoring, and materials testing.

The Agilent 6010A is a high-performance spectrometer known for its precision and versatility. It utilizes advanced optical technologies to provide exceptional wavelength accuracy and resolution. This model is particularly useful in laboratories where reliable data is critical, offering a wide spectral range and effective noise reduction features. Its user-friendly interface simplifies complex analyses, making it suitable for both seasoned professionals and newcomers.

Following closely, the Agilent 6011A is recognized for its robust capabilities in laboratory environments. This device incorporates advanced signal processing techniques, enabling high-throughput measurements without compromising on quality. The 6011A is ideal for real-time monitoring applications, ensuring that users can make informed decisions based on accurate, timely data. Its comprehensive software suite is designed to enhance data analysis, allowing for seamless integration with existing laboratory workflows.

The 6012B variant enhances the functionality further by introducing additional features tailored for specific applications. With a focus on flexibility, the 6012B supports multiple measurement modes, including direct and differential detection. This model excels in complex measurements, allowing for greater analytical depth and insights. The built-in calibration options ensure consistent performance, making it a reliable choice for various research and development tasks.

Lastly, the Agilent 6015A model stands out with its leading-edge technology, designed for the most demanding applications. It boasts enhanced sensitivity and an improved dynamic range, making it perfect for trace analysis in challenging environmental samples. The 6015A’s advanced reporting tools provide detailed analytics, helping scientists and researchers interpret results efficiently. Its compact design also makes it suitable for laboratory spaces with limited room, without sacrificing performance.

Together, these models showcase Agilent Technologies' commitment to delivering high-quality, innovative solutions that empower users to achieve their analytical goals effectively and efficiently. Whether in a research, clinical, or industrial setting, the 6010A, 6011A, 6012B, and 6015A continue to set standards in precision instrumentation.