Agilent Technologies 60501B Performance Test Record Agilent 60502B Load Module Page 2

Page 18

PERFORMANCE TEST RECORD - Agilent 60502B LOAD MODULE (Page 2 of 2)

 

Model Agilent 60502B

Report No.____________________

Date______________________________

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Test Description

Minimum

Results

 

Maximum

Measurement

 

 

Specification

 

 

Specification

Uncertainty

60 Ampere Range Programming

CONSTANT CURRENT MODE TESTS

 

 

 

 

 

 

 

 

 

and Readback

 

 

 

 

 

 

High Current (60A)

59.865

____________A

60.135

25mA

Front Panel Display

AOUT -0.095

____________A

AOUT +0.095

25mA

Low Current (1A)

0.924

____________A

1.076

427∝A

Front Panel Display

AOUT -0.065

____________A

AOUT +0.065

427∝A

6 Ampere Range Programming

 

 

 

 

 

 

and Readback

 

 

 

 

 

 

High Current (6A)

5.919

____________A

6.081

2.4mA

Front Panel Display

AOUT -0.068

____________A

AOUT +0.068

2.4mA

Low Current (1A)

0.924

____________A

1.076

427∝A

Front Panel Display

AOUT -0.065

____________A

AOUT +0.065

427∝A

Voltage Programming and

CONSTANT VOLTAGE MODE TESTS

 

 

 

 

 

 

 

 

 

Readback

 

 

 

 

 

 

High Voltage (60V)

59.890

____________V

60.110

845∝V

Front Panel Display

VOUT -0.075

____________V

VOUT +0.075

845∝V

Low Voltage (3V)

2.947

____________V

3.053

35∝V

Front Panel Display

VOUT -0.046

____________V

VOUT +0.046

35∝V

Low Resistance Range

CONSTANT RESISTANCE MODE TESTS

 

 

0.984

 

 

1.016

 

 

Resistance (1Ω)

____________Ω

 

 

Resistance (0.05Ω)

0.0416

____________Ω

0.0584

 

 

Middle Resistance Range

24.1

 

 

39.6

 

 

Resistance (30Ω)

____________Ω

 

 

Resistance (1Ω)

0.989

____________Ω

1.011

 

 

High Resistance Range

61.1

 

 

3243

 

 

Resistance (120Ω)

____________Ω

 

 

Resistance (12Ω)

10.9

____________Ω

13.3

 

 

 

 

TRANSIENT SLEW TEST

 

 

 

Fast Slew Transient

75

 

 

125

 

 

Slew Rate 0.05 A/∝s

____________∝s

 

 

Slew Rate 2.5 A/ms

1.5

____________ms

2.5

 

 

 

 

CONSTANT CURRENT PARD TEST

 

 

 

Current (10A)

0

__________mA

4mA RMS

 

 

 

 

 

 

 

 

 

 

18

Image 18
Contents Electronic Load Modules Certification Safety Summary Table of Contents Troubleshooting Precautions About This ManualManual Revisions Module Initialization Store calibration constants in EepromIntroduction Test Equipment RequiredCC Mode Test 60501B 60502BCV Mode Test Test Setup BCR Mode Test Check the middle ohms range as follows Press Then Transient Operation and Slew Circuit Test CC Mode Pard Test CC Mode Power LimitPage Performance Test Record Agilent 60501B Load Module Page 1 Test Description Minimum Results Specification UncertaintyPerformance Test Record Agilent 60501B Load Module Page 2 MeasurementPerformance Test Record Agilent 60502B Load Module Page 1 Agilent 60502BPerformance Test Record Agilent 60502B Load Module Page 2 Model Agilent 60502BHow To Order Parts Reference DesignatorsPart Description Abbreviations Designation Number DescriptionFUSE-SUBMINIATURE .5A Designation Number IC Quad Nand SET/RESET Latch Cmos Model Agilent Part NumberNot Used RESISTOR-NET 10-PIN SIP MULTI-VALUE Model RT2 Agilent 60501B/60502B Mechanical Parts Quantity Description NumberPage Schematic Diagrams Schematic Diagram NotesCommon +5V Signal-Name Descriptions Mnemonic Function Sheet1D, 1C Mnemonic Sheet 1C, 7D 1B, 6D8C,8B Signal-Name Descriptions Mnemonic Function Sheet Power BoardPage Page Page Page Page Page
Related manuals
Manual 11 pages 42.07 Kb Manual 11 pages 45.05 Kb

60501B, 60502B specifications

Agilent Technologies 60502B is a high-performance signal generator designed for a wide range of applications, including aerospace, telecommunications, and research and development. As part of Agilent's advanced suite of test and measurement solutions, the 60502B provides users with the capability to generate high-quality signals with exceptional accuracy and stability.

One of the standout features of the Agilent 60502B is its frequency range, which spans from 9 kHz to 6 GHz. This broad range enables engineers and technicians to work with various modulation formats and signal types, making it highly versatile for testing and development. The generator supports continuous wave (CW) signals as well as complex modulated signals, accommodating both narrowband and broadband applications.

The 60502B employs advanced direct digital synthesis (DDS) technology, ensuring accurate frequency generation with low phase noise. This technology is essential for applications that require high precision and minimal signal distortion, such as radar and communication systems. The generator’s phase noise performance is exceptional, which enhances its capability for demanding applications requiring signal integrity.

Another key characteristic of the Agilent 60502B is its output power handling. It can deliver output levels ranging from -130 dBm up to +10 dBm, enabling users to drive various loads and achieve optimal testing conditions. This flexibility is particularly beneficial in laboratory environments where different power levels are needed for different devices under test.

The user interface of the 60502B is designed for efficiency and ease of use. It incorporates a large display that provides real-time feedback on settings and signal characteristics, facilitating quick adjustments during testing. Furthermore, the generator is equipped with USB and GPIB interfaces, enabling seamless integration into automated test systems and easy connectivity to computers for remote control and data logging.

In addition to its robust hardware features, the Agilent 60502B supports a variety of modulation schemes, including AM, FM, PM, and pulse modulation. This capability allows users to simulate real-world signals and assess the performance of receivers and other components under various conditions.

Overall, the Agilent Technologies 60502B signal generator stands out in terms of performance, versatility, and user-friendliness. Its wide frequency range, advanced DDS technology, significant power output capabilities, and support for multiple modulation types make it an indispensable tool for engineers across various industries engaged in testing and development activities. Whether for research projects or routine maintenance tasks, the 60502B is built to deliver reliable and precise results.