TROUBLE | PROBABLE CAUSE | CORRECTIVE ACTION |
Image is too dark | Imaging plate has been | Adjust brightness with software. If this is |
| overexposed | not possible, retake image at lower |
|
| exposure. See exposure guidelines |
|
| Section 5.1. |
|
|
|
Image contains ghost | Imaging plate was not | Increase the amount of time that is used |
images or shadows. | completely erased. | to erase plates. Alternatively, increase the |
|
| intensity of the light source. To te st if an |
|
| imaging plate is completely erased, |
|
| simply mount an erased plate on a |
|
| carousel and scan it. If no image is |
|
| obtained, your procedure is effective at |
|
| completely erasing imaging plates. |
Imaging plate was exposed with the back of the IP facing the
Insure the imaging plates are inserted properly into the protective envelope and the proper orientation to the
Imaging plates have been stored in protective envelopes for too long a period.
Do not store imaging plates in protective envelopes for more than one week.
Partial erasure of the image due to exposure to light during handling of the imaging plates.
Do not leave exposed imaging plates in
Image shows artifacts. | The imaging plate surface is |
| not clean and has dirt or stains |
| on it. The surface may be |
| scratched. |
| The imaging plate was |
| removed from the cassette too |
| quickly, which resulted in a |
| static discharge. |
| A cosmic event occurred and |
| deeply exposed the image |
| plates with cosmic rays. This |
| may occur during shipment. |
Clean the imaging plate as outlined in Section 7.1. If the plate is scratched or stained, do not use the plate again.
Clean the imaging plate with anhydrous isopropyl alcohol AND remove from cassette more slowly.
Erase the image plate and rescan until the artifacts are no longer present.
Imaging plates fall out into the EPIX SCANNER carousel well
The imaging plates have not been loaded properly. See Section 5.3 for more information.
Practice loading erased imaging plates in full light until you become familiar with the loading procedure. You will hear a “click” when the panoramic screen is loaded properly.
The red light stays on and | The lid to the carousel well has | Do not raise the lid while the unit is |
the scanner stops | been opened. | scanning. |
| Hardware failure | Call your service representative |
|
|
|
Logos Imaging EPIX Scanner |
| 21 |
User Manual |
|
|