Epson EPIX manual Troubleshooting, Trouble Probable Cause Corrective Action

Page 21

7.3. TROUBLESHOOTING

TROUBLE

PROBABLE CAUSE

CORRECTIVE ACTION

No power

Scanner not plugged in

Check power cord at outlet and at the

No green light

 

connection on the back of the scanner.

 

 

Turn on the scanner and then turn on the

 

 

computer and monitor.

 

 

 

 

Blown fuse

See Section 7.2

 

Main power switch or power

Call your service representative.

 

supply is bad

 

 

 

 

 

Outlet does not have power

Ensure that the outlet is grounded and

 

 

has power.

 

 

 

Red, Green and/or Yellow

Defective LED

Call your service representative

LED(s) do not work;

 

 

system operates normally

 

 

The Scanner does not initialize when the software is opened

The scanner has not been

If the green power indicator light is off,

turned on

switch on the EPIX SCANNER per

 

Section 5.5.

 

 

The lid on the carousel well is

Close the lid.

open

 

 

The USB cable between the

Reconnect the cable. Check for tightness.

 

scanner and computer is loose

Swap with a known good cable if

 

or defective.

possible.

 

 

 

 

The computer does not

Turn off the scanner and re-install the

 

recognize that the scanner is

EPIX SCANNER systems software

 

connected or the scanner has

driver (see section 4.3).

 

not been added to the Device

Turn the unit on.

 

Manager

 

 

 

 

 

There is a hardware problem

Contact your service representative

 

with the EPIX SCANNER.

 

 

 

 

After scanning, no image

There are no imaging plates on

Ensure that the exposed imaging plate is

appears on the monitor

the carousel.

properly mounted for scanning

 

 

 

 

The imaging plate is mounted

The imaging plates are to be mounted

 

backwards on the carousel.

with the white or blue surface visible.

 

 

Turn incorrectly mounted plates around

 

 

and rescan

The imaging plate was erased prior to scanning.

Ensure that imaging plates are mounted under low light conditions as outlined in Section 4.1.

Hardware failure.

Call your service representative.

X-ray source failed.

Call your service representative.

Logos Imaging EPIX Scanner

20

User Manual

 

Image 21
Contents EPIXScanner USB Supplies and Replacement Parts Epix Digital Imaging System by Logos ImagingTechnical Support Table of Contents Epix Scanner is not intended for medical imaging of humans USEImportant Information Safety RequirementsDo not leave unscanned, exposed imaging plates in light PRE-INSTALLATION Information Recommendations Abbreviations and Explanation of SymbolsPurpose of the Manual Quantity Description Physical DescriptionCustomer Supplied Components Optional Equipment from Logos ImagingScanner Front View Epix Scanner Laser ScannerLogos Imaging Epix Scanner Minimum Computer Requirements Recommended Computer RequirementsElectrical Requirements Compliance to StandardsSite Selection InstallationUnpacking the Unit Cable Power Hardware Setup and ConnectionsLoading the Epix Driver Software USBSystem Operating Instructions Preparing Your Current X-RAY EquipmentErase the Imaging Plate 10 in shown. Fig Taking AN X-RAYTurning OFF the Epix System Scanning the Imaging PlateSoftware Operating Instructions Cleaning the System Maintenance ProceduresCleaning the Scanner Cleaning Imaging PlatesScanner Maintenance Operator MaintenanceCarousel Maintenance Imaging Plate MaintenanceTrouble Probable Cause Corrective Action TroubleshootingQuickly, which resulted in a Disposal of Waste Materials and Inoperartive Parts Material ItemsComments Storage and Shipment StorageShipment Specifications Technical SpecificationsEpix Scanner Epix Imaging PlatesWarranty Statement How Storage Phosphor Imaging Plates Work Appendix a Storage Phosphor TechnologySunlight Incandescent LightingPractical Test Inkjet printers Appendix C Optional PrinterDye Sublimation printers Application softwarePlease return this registration to Appendix D Warranty Registration