Teledyne IGE 3000 instruction manual Quick Start Information, To Turn on the TUBE’S Filaments

Page 5

1. Quick Start Information

Apply Power

Apply 24 VDC Power to the “D” connector Pin 3 and 4.

After power is applied the IGE 3000 will go through self test, CPU initialization, and LED test. (The individual LEDs will sequentially illuminate then the IGE 3000 enters the “No Mode” and is operational).

Operate Tube

Before operating tube, ensure vacuum is below 1x10-2 Torr. The IGE 3000 has an automatic shutoff feature that will not allow filament turn on at pressures above 1x10-2 Torr. The shutoff feature will also automatically turn off the filament if it is turned on (below 1x10-2 Torr) and pressure rises above 1x10-2

TO TURN ON THE TUBE’S FILAMENTS:

“DOUBLE CLICK” the select button. (note: Double clicking the SELECT button while the filament is already On will turn it OFF).

Initially dashes will remain on the display. However, as the tube heats up and starts operating, within seconds the pressure will be displayed. LED’s (F1 or F2) will be illuminated, indicating which filament is selected and that it is turned on.

High Low

F1 F2

Emis DGas SP K Fil

2X=On/Off

169-082010_Ion-Gauge-3000

Page 5 of 21

Image 5
Contents Teledyne Hastings Manual Print History Table of Contents 169-082010Ion-Gauge-3000 To Turn on the TUBE’S Filaments Quick Start InformationGeneral Information SpecificationsGeneral Information IGE-3000 Mini Bayard/Alpert B/A TubeMaterials exposed to Vacuum Installation InstallationReceiving Inspection MountingB/A Tube Installation Connector Electrical Pin outsMounting B/A tube to IGE-3000 Electronics Module EnvironmentalOperation Apply PowerOperation Front Panel OverviewOperating Modes Factor Mode DGas ModeFil filament Mode SP set point ModeRemote Filament Control Bayard Alpert B/A Tube Theory of OperationAdvance Setup Guide Set Up GuideRS485 Address Baud RateRestore Eeprom Turnaround DelayData Bits/Parity Bits/Stop Bits Save EepromRS232/RS485 Wiring for IGE-3000 RS232/485 Wiring for RS232/485RS 232/RS ION Gauge Device Status 4-DIGIT Octal Word Command SyntaxDevice Status Command Description Format Sample Response Interrogation CommandsParameter Modification Commands Non-Warranty Repair Policy Warranty Repair PolicyDrawings & Schematics