8. Device Status

8.1.ION GAUGE DEVICE STATUS: 4-DIGIT OCTAL WORD

When requested to transmit its status the IGE-3000 responds with a 4-digit octal word.

FIGURE 7 -1 – DIGIT OCTAL WORD

bit 11: Serial Receiver Overload

bit 10: Main Board EEPROM Error.

bit 9: Communication Syntax Error.

bit 7 Emission Setting.

bit 8 Emission Setting.

bit 6: Filament Selection. (0=filament 1, 1=filament 2)

FIGURE 8- BITS 7 & 8 EMISSION CODE

BITS 7 & 8

BIT 8

BIT 7

NO EMISSIONS

0

0

0.01 Ma

0

1

0.1 mA

1

0

1.0 mA

1

1

bit 5: Filament Status. (0=non-operating, 1=operating)

bit 4: Degas (0=degas off, 1=degas on)

bit 3: Pressure less than (0.1e-9 Torr) Set if true.

bit 2: Pressure Greater than (1.0e-3 Torr) set if true.

bit 1: High Set point Alarm Set if pressure is greater than the high set point

bit 0: Low Set point Alarm Set if pressure is less than the low set point

8.2.Command Syntax

In the following examples of syntax codes, the special characters are explained:

The characters in square brackets [ ] represents a command string, either upper or lower case command characters accepted. All characters must follow each other in the string with no spaces or other characters.

The characters within wavy brackets { } contain choices for the appropriate command.

The characters within the symbols < > are the common abbreviations for the one digit ASCII control codes which they represent, (e.g. <CR> represents carriage return).

169-082010_Ion-Gauge-3000

Page 18 of 21

Image 18
Contents Teledyne Hastings Manual Print History Table of Contents 169-082010Ion-Gauge-3000 Quick Start Information To Turn on the TUBE’S FilamentsGeneral Information SpecificationsGeneral Information IGE-3000 Mini Bayard/Alpert B/A TubeMaterials exposed to Vacuum Receiving Inspection InstallationInstallation MountingMounting B/A tube to IGE-3000 Electronics Module Connector Electrical Pin outsB/A Tube Installation EnvironmentalOperation Apply PowerOperation Front Panel OverviewOperating Modes Fil filament Mode DGas ModeFactor Mode SP set point ModeRemote Filament Control Theory of Operation Bayard Alpert B/A TubeRS485 Address Set Up GuideAdvance Setup Guide Baud RateData Bits/Parity Bits/Stop Bits Turnaround DelayRestore Eeprom Save EepromWiring for IGE-3000 RS232/485 Wiring for RS232/485 RS 232/RSRS232/RS485 Command Syntax Device StatusION Gauge Device Status 4-DIGIT Octal Word Interrogation Commands Parameter Modification CommandsCommand Description Format Sample Response Warranty Repair Policy Non-Warranty Repair PolicyDrawings & Schematics
Related manuals
Manual 21 pages 23.15 Kb