Teledyne IGE 3000 RS 232/RS, RS232/RS485, Wiring for IGE-3000 RS232/485 Wiring for RS232/485

Page 17

7. RS 232/RS 485

7.1.RS232/RS485

The IGE-3000 is designed to communicate with external devices by either RS232 or RS485 via the two RJ-11 connectors. Refer to section 3.2 Figure 2 for connector locations. RS232 via the left connector on and the RS485 is via either connector. Software capable of communicating with the IGE-3000 is available upon request.

Refer to Figure 7 below for pin out and wiring information.

7.1.1.Wiring for IGE-3000 RS232/485

7.1.2.Wiring for RS232/485

 

RS 485 half duplex

 

 

RS 485 full duplex

Pin #1

T+/R+

Pin #1

 

T+

 

 

 

 

 

 

Pin #2

 

T-/R-

Pin #2

 

T-

Pin #3

GND

Pin #3

 

GND

 

 

 

 

 

 

Pin #4

GND

Pin #4

 

GND

 

 

 

 

 

 

Pin #5

n/a

Pin #5

 

R-

 

 

 

 

 

 

Pin #6

 

n/a

Pin #6

 

R+

169-082010_Ion-Gauge-3000

Page 17 of 21

Image 17
Contents Teledyne Hastings Manual Print History Table of Contents 169-082010Ion-Gauge-3000 To Turn on the TUBE’S Filaments Quick Start InformationGeneral Information SpecificationsGeneral Information IGE-3000 Mini Bayard/Alpert B/A TubeMaterials exposed to Vacuum Installation InstallationReceiving Inspection MountingB/A Tube Installation Connector Electrical Pin outsMounting B/A tube to IGE-3000 Electronics Module EnvironmentalOperation Apply PowerOperation Front Panel OverviewOperating Modes Factor Mode DGas ModeFil filament Mode SP set point ModeRemote Filament Control Bayard Alpert B/A Tube Theory of OperationAdvance Setup Guide Set Up GuideRS485 Address Baud RateRestore Eeprom Turnaround DelayData Bits/Parity Bits/Stop Bits Save EepromRS232/RS485 Wiring for IGE-3000 RS232/485 Wiring for RS232/485RS 232/RS ION Gauge Device Status 4-DIGIT Octal Word Command SyntaxDevice Status Command Description Format Sample Response Interrogation CommandsParameter Modification Commands Non-Warranty Repair Policy Warranty Repair PolicyDrawings & Schematics
Related manuals
Manual 21 pages 23.15 Kb