Troubleshooting
If your cleaner does not operate satisfactorily, please check the tables below for possible causes before calling your authorized service center.
WARNING
High voltage inside - dangerous shock hazard.
DO NOT attempt to disassemble or repair the cleaner.
Problem | Cause | What to do |
Cleaner | Cleaner not plugged in | Plug into functioning |
will not | properly. | electrical outlet. |
start. |
| . |
|
| |
| Circuit board fuse blown | Call nearest authorized |
|
| service center. |
Cleaner | Heater malfunctions. | Call nearest authorized |
operates |
| service center. |
but does |
|
|
not heat |
|
|
solution |
|
|
Decreased | Solution is not degassed. | Make sure that tank was |
ultrasonic |
| filled with warm tap water |
activity. |
| plus cleaning chemistry and |
|
| has run |
| Solution is spent. | Change solution. |
| Solution level is incorrect | Adjust solution +/- 3/8 inch |
| for load. | from current level. |
| Tank bottom is covered with | Empty, then clean tank with |
| soil particles. | warm water. Wipe with a |
| Using deionized water in | nonabrasive cloth. |
| Deionized water does not | |
| the tank. | cavitate as actively as soapy |
|
| tap water. |
13 Troubleshooting