Performance

Check your cleaner periodically to test the level of activity of the ultrasonic cavitation. Frequency of testing will depend on your use of the cleaner, however, we suggest running this test monthly.

Glass Slide Test

You will need the following equipment:

Frosted microscope glass slide (1” x 3”), such as ESCO #2951F, or equivalent;

No. 2 lead pencil; and

General purpose dish detergent.

Test procedure:

1.Prepare a fresh solution with general purpose dish detergent (concentration 1%) and warm tap water (120º - 140ºF).

2.Fill the cleaner to within one(1) inch of the tank top.

3.Turn the ultrasonics on for at least five minutes to allow for degassing.

4.Prepare the glass slide by first wetting the frosted portion with tap water.

5.With the No. 2 pencil, on the frosted portion make an “X” from corner to corner.

6.Immerse the frosted end of the slide into the solution. Hold the slide vertically and center it in the solution.

7.Turn ultrasonic switch to “On”.

14 Troubleshooting

Image 18
Contents Operator’s Manual Warranty Contents Page Safety Precautions To avoid electrical shockTo prevent personal and/or property damage To prevent damage to the cleaner Introduction PC620 General InformationAccessories For Your Cleaner Unpacking Your CleanerInstalling Your Cleaner Equipment Specifications Model Tank Overall Weight Heater Input Volume Size PowerHow Ultrasonic Cleaning Works Operating Your Cleaner Before You Start CleaningCleaning Items Getting ReadyOptimizing Your Cleaner Tanks TemperatureApplication Hints Cleaning Solutions Solution Types Alkaline strength RemovesChemistry Concentrations Chemicals Harmful to Your TankTroubleshooting Performance Glass Slide TestService Authorized Service CentersCPN-214-159