OPERATING INSTRUCTIONS

Timer function

The timer function has two modes as follows:

1.Normal mode: starts immediately after start of the running

2.Timer mode 1: starts when the chamber temperature is 2.0°C lower than the set point. When conducting the sterilization, always select "Timer mode 1".

On timer mode 1, the timer starts when the chamber temperature is 2.0°C lower than the set point. After timer starting, the temperature indicator blinks, alarm lamp lights, and intermittent tone sounds if the chamber temperature is 5.0°C or more lower than set point due to the door opening, etc.

In spite of the alarm condition, the unit continues to run. The timer is reset to initial setting when the chamber temperature is 2.0°C lower than the set point again.

Set point

Start

Normal mode

Timer mode 1 (For sterilization)

Sterilization

Refer to the table below when setting the sterilization temperature and duration.

 

Sterilization temperature

Sterilization duration

 

 

135to145°C

3 to 5 hours

 

 

160to170°C

2 to 4 hours

 

 

180to200°C

0.5 to 1 hour

 

In case of sterilizing a lot of samples, extend the sterilization time for more than an hour.

When sterilizing the combustibles, such as cotton capped tube and instrument which are covered with paper, do not put them on the shelves directly. Use the wire basket and upper shelves instead.

When sterilizing the pipettes or the glass pipes, waterdrop must be dried out clearly, otherwise sterilization will not be completed.

Take out the articles in the chamber when the chamber temperature is lower than 50°C.

Note:

Do not open the door during sterilization.

Do not open the door when the chamber temperature is higher than 100°C.

Image 14
Contents MOV-112S MOV-212S Contents Precautions for Safe Operation Precautions for Safe Operation \WARNING Precautions for Safe Operation Environmental Conditions Sterilizer Components Sterilizer Components Control panel and keypad Location away from heat generating sources InstallationInstallation site Location with a sturdy and level floorPrerunning InstallationBefore Commencing Operation Basic operating flow chart of control panel Operating InstructionsTimer function SterilizationStable running Chamber temperature 50C, Delay time 1 hour Stable running with delay timeStart of operation with no change of setting Press the power keySterilization with timer mode Alarms & Safety Functions Cleaning the inside Routine MaintenanceCleaning of unit Cleaning the frameIf the alarm function operates Trouble ShootingKey operation is disabled Disposal of UnitSpecifications PerformanceSafety check sheet