SPECIFICATIONS

Name

Sterilizer

 

 

 

 

Model

MOV-112S

 

MOV-212S

 

 

 

 

External dimensions

W580 x D595 x H820 (mm)

 

W730 x D645 x H870 (mm)

 

 

 

 

Internal dimensions

W450 x D450x H450 (mm)

 

W600 x D500x H500 (mm)

 

 

 

 

Effective capacity

90 L

 

150 L

 

 

 

Exterior

Electrically galvanized steel plate, Acrylic resin painted

 

 

Interior

Stainless steel plate (SUS 304)

 

 

Door

Electrically galvanized steel plate, Acrylic resin painted

 

 

Observation window

Reinforced glass x 3, W150 x H380 mm

 

 

 

Insulation

Glass wool

 

 

 

Rack

Stainless steel plate, Stainless wire (SUS 304)

 

 

Temperature controller

Sensor K, PID control

 

 

 

Temperature display

Digital display

 

 

 

Timer

Electronic timer with delay timer

 

 

 

Circuit breaker

15A

 

 

 

 

 

Overheat prevention

Built-in thermister (electric circuit), Thermal guard

mechanism

 

 

 

Heater

1.1 kW

 

1.2 kW

Fan motor

Output 10 W

 

 

 

 

Accessories

2 racks, 4 rack supports

 

3 racks, 6 rack supports

Weight

50 kg

 

66 kg

 

 

 

 

Note: Design or specifications will be subject to change without notice.

 

PERFORMANCE

Model

 

MOV-112S

MOV-212S

Usable environment condition

 

 

 

Temperature; 0°C to 35°C

 

Extent chamber temperature

 

 

 

Ambient temperature +5°C to 200°C

 

Temperature control range

 

 

 

±4°C

 

Time of reaching maximum

 

60 minutes

70 minutes

Temperature (0°C 200°C)

 

 

 

 

 

 

 

 

AC110V

60Hz, 1030W

 

 

Power source, Rated power consumption

AC115to120V

60Hz, 1130W

AC110 to 120V

60Hz, 1230W

AC220

50/60Hz, 1130W

AC220V 50/60HZ, 1230W

 

 

AC230 to 240V

50Hz, 1130W

AC230 to 240V

50Hz, 1230W

 

 

 

 

 

 

 

Note: The unit with CE mark complies with EC directives 89/336/EEC, 93/68/EEC and 73/23/EEC.

Image 21
Contents MOV-112S MOV-212S Contents Precautions for Safe Operation Precautions for Safe Operation \WARNING Precautions for Safe Operation Environmental Conditions Sterilizer Components Sterilizer Components Control panel and keypad Installation site InstallationLocation away from heat generating sources Location with a sturdy and level floorInstallation Before Commencing OperationPrerunning Operating Instructions Basic operating flow chart of control panelSterilization Timer functionStable running Start of operation with no change of setting Stable running with delay timeChamber temperature 50C, Delay time 1 hour Press the power keySterilization with timer mode Alarms & Safety Functions Cleaning of unit Routine MaintenanceCleaning the inside Cleaning the frameKey operation is disabled Trouble ShootingIf the alarm function operates Disposal of UnitPerformance SpecificationsSafety check sheet